USB Interface Communication

:

The 

9825  indicator  can  connect  to  a  PC  via  USB  cable.  First  a  USB  driver  must  be  installed  on  the 

PC  to  access  the  9825.  Measurement  data  can  be  accessed  by  using  a  terminal  emulation  application 

such  as  HyperTerminal.

 

The  USB  Port  output  has 

Two

  fixed  strings: 

ASCII and Condec.

Analog Output Interface

:

Analog Output Calibration:

The mode of the Analog Output can be selected from its Output Type sub-menu. There are four

modes of analog output: 4-20mA, 0-24mA, 0-5V and 0-10V. 

Please refer to the wiring section for

the proper jumper setting of the optional analog output board. 

To calibrate the Analog Output, follow

the steps below:

Scale Output:

1.

While within the Analog Output menu, scroll to Scale Output and press

 the

 

↓ (Pk/Val)

button to initiate the Scale Output sequence.

2.

Scale Output is set by inputting a Low and High force value. To set a given value, use

the numerical field on the screen to input the desired force. The first character can be

used to switch the sign convention from + to - and back. Press the ↓ (Pk/Val) button to

save the setting.

Fine Tune:

Before performing this part of the device setup, the Analog Output of the 9825 should be

connected to whatever instrument will be accepting and measuring the analog signal.

1.

In the Analog Output setup menu, scroll to Fine Tune and press the ↓ (Pk/Val) button to

initiate the Fine Tune sequence.

2.

The screen will display "0%", which indicates the lowest point of the analog scale. For

voltage outputs, this is 0VDC. For current outputs this is either 0mA (0-24mA) or 4mA

(4-20mA).

3.

By adjusting the numerical value on the screen, the Analog output will be Fine Tuned.

The digit furthest to the left creates the greatest change in output, while the digit furthest

to the right creates the smallest change in output. Adjust this number until the measured

value on the connected meter or PLC shows the minimum point on the analog scale.

Press Menu (Enter) to save this value and proceed.

4.

Repeat this process for the 50% point. For a 0-5V setting the output will be 2.5V. For a

4-20mA setting the output will be 12mA and so on.

5.

Repeat this process for the 100% point.

Notes:

 

- The analog output mode setup to 4mA-20mA: If the load  0kg, the voltage output is

0. If the load is  the full range of the scale, than the voltage output is 24mA.

- The analog output mode setup to 0-10V: If the load  0kg, the voltage output is 0. If

the load is  the full range of the scale, than the voltage output is 10.8V.

Interface Inc. www.interfaceforce.com

12

9825 Installation & User Manual

Содержание 9825

Страница 1: ...9825 Installation User Manual Installation User Manual 9825...

Страница 2: ...instructions Warning Warning means potential danger Example property damage personal Injury or even death Pre Installation Warnings Warning This device must be installed and connected by a profession...

Страница 3: ...USB Interface Communication Analog Output Interface I O Control Applications Setpoint Application Alarm Application Indicator Info Software Version Appendix Appendix 1 Command Output Format 1 Continuo...

Страница 4: ...is 20 C to 65 C 4 F to 149 F working environment temperature is 10 C to 104 F 14 F to 104 F relative humidity no more than 95 Non Condensing The 9825 digital indicator uses panel installation which re...

Страница 5: ...load cell signal connection This indicator provides a 4 5 volt DC excitation voltage to the load cell s The voltage difference between SIG and SIG is about 0 9mV when connected to a load cell with a 2...

Страница 6: ...Input Terminal Pin Assignment Four Wire Analog Load Cell or Junction Box Connection Six Wire Analog Load Cell or Junction Box Connection Interface Inc www interfaceforce com 3 9825 Installation User...

Страница 7: ...tput for current output 4 20mA 0 24mA or voltage output 0 10V 0 5V Please note that voltage and current outputs cannot be used at the same time We suggest using a PLC or PC to monitor the analog outpu...

Страница 8: ...l can be connected to the positive or negative of the power supply The maximum power output of each relay is 90W 5A Output control interfaces and load connection diagram Output control interfaces and...

Страница 9: ...ce value will be displayed Display Details The 9825 utilizes a 128 x 32 dot OLED display with adjustable LED back light The table below summarizes the display annunciators Basic Force Display Force Di...

Страница 10: ...ay mode PK Val Function Cycle among real time peak and valley display modes When in the setup menu Function Enter sub menu Decrease a value when used as a directional key Reset When in display mode Re...

Страница 11: ...mooths signal by averaging samples over a given span 1 Integer values from 1 to 100 Units Cal Basic Unit Select engineering units displayed lbf lbf ozf N kN g kg UNIT_NONE Nominal Capacity Sets displa...

Страница 12: ...8 bit None 8 bit Even 7 bit Even 7 bit Odd Port Mode Set port mode Demand Demand Continuous Protocol Set Protocol see protocol descriptions in the appendix Condec Condec ASCII Analog Out Output Type S...

Страница 13: ...he screen Press the Pk Val button to initiate the calibration process 4 The term Fixture will show on the screen At this point the load cell should be set into its fixturing but with no additional cal...

Страница 14: ...follow the steps below 1 Press and hold the Menu button to enter the setup menu A beep will sound as the setup menu is activated 2 Using the Menu button scroll until Calibration is displayed on the s...

Страница 15: ...t of the 9825 should be connected to whatever instrument will be accepting and measuring the analog signal 1 In the Analog Output setup menu scroll to Fine Tune and press the Pk Val button to initiate...

Страница 16: ...s less than the value of Input Point 4 but greater than the value of Input Point 3 The symbol will show on the display The OUT 4 relay will close Otherwise the symbol will show on the display and the...

Страница 17: ...relay will open 4 When the load is less than the value of Input ExtraLow The symbol will show on the display The alarm will sound The OUT 4 relay will close The display will send a warning message Oth...

Страница 18: ...Format 1 Continuous Mode ASCII In this mode of communication the indicator transmits the data frame continuously The load value in the frame is expressed in ASCII Interface Inc www interfaceforce com...

Страница 19: ...device PC will output demand command through the serial ports when the scale is in the normal loading status The demand command format is shown below The serial output data format is as follows Interf...

Страница 20: ...ndec Demand Output Demand Commands P Print T Tare Z Zero G Gross N Net Condec Continuous Output Note The starting address 40001 of MODBUS is not suitable for SIEMENS soft Interface Inc www interfacefo...

Страница 21: ...C 10 to 90 non condensing at F 10 to 90 non condensing POWER Supply VDC 24 VDC with supplied 120V 60Hz AC DC adapter or 9 36 VDC external supply Power Consumption W 6 RMS 8 Peak MECHANICAL Dimensions...

Страница 22: ...freight insurance and proper packaging to prevent breakage in transit Interface warranty does not apply to defects resulting from action of the buyer such as mishandling improper interfacing operatio...

Отзывы: