Calibration Live or Key-In
Set calibration type Live
Live, Key-In
Set Pos Span
Set span from zero
to positive capacity
Press
↓
and enter to start
sequence
Fixture
Press
↓
to start sequence
Set Neg Span
Set span from zero
to negative capacity
Set Zero Point
Set zero
Cal Stability
A greater value can
produce higher
accuracy calibration
points but will
require a more
stable mV/V input
signal during
calibration as well.
1
Integer values from 0 to 320
represent the number of
samples averaged when
capturing a calibration point.
Larger values = greater stability
required
USB Setup Baud Rate
Serial
communication
rate in bits per
second
9600
2400, 4800, 9600, 19200, 38400,
57600,115200
Bit / Parity
Set Binary format
and check bit
8-bit
None
8-bit None, 8-bit Even, 7-bit
Even, 7-bit Odd
Port Mode
Set port mode
Demand
Demand, Continuous
Protocol
Set Protocol (see
protocol
descriptions in the
appendix)
Condec
Condec, ASCII
Analog Out Output Type
Set Analog output
type
0-10V
4-20mA, 0-10V, 0-5V, 0-24mA
Scale Output
Adjust Low and High points
using keypad
Fine Tune
Adjust 0%, 50%, 100% output
points using keypad
Relay
IO Application
None
None, Setpoint, Alarm
Input Point n
(Setpoint)
4000
Adjust values using keypad
Input Hysteresis n
(Setpoint)
200
Input High Point
(Alarm)
5000
Input Low Point
(Alarm)
3000
Custom In-1
None
None, Reset Key, Tare Key, Print
Key
System
Version / Date
Display firmware
version and date
Press
↓
to view
Unique ID
Display Unique ID
Power-On Tare
Disable
Off, On
Sys Reset
Reset to default
settings.
Press
↓
to execute
Advanced Menu
Password Required
Enter Password
336699
to
access the Advanced Menu
Interface Inc. www.interfaceforce.com
9
9825 Installation & User Manual