Installation & User Manual

Installation & User Manual

9

8

2

5

Содержание 9825

Страница 1: ...9825 Installation User Manual Installation User Manual 9825...

Страница 2: ...instructions Warning Warning means potential danger Example property damage personal Injury or even death Pre Installation Warnings Warning This device must be installed and connected by a profession...

Страница 3: ...USB Interface Communication Analog Output Interface I O Control Applications Setpoint Application Alarm Application Indicator Info Software Version Appendix Appendix 1 Command Output Format 1 Continuo...

Страница 4: ...is 20 C to 65 C 4 F to 149 F working environment temperature is 10 C to 104 F 14 F to 104 F relative humidity no more than 95 Non Condensing The 9825 digital indicator uses panel installation which re...

Страница 5: ...load cell signal connection This indicator provides a 4 5 volt DC excitation voltage to the load cell s The voltage difference between SIG and SIG is about 0 9mV when connected to a load cell with a 2...

Страница 6: ...Input Terminal Pin Assignment Four Wire Analog Load Cell or Junction Box Connection Six Wire Analog Load Cell or Junction Box Connection Interface Inc www interfaceforce com 3 9825 Installation User...

Страница 7: ...tput for current output 4 20mA 0 24mA or voltage output 0 10V 0 5V Please note that voltage and current outputs cannot be used at the same time We suggest using a PLC or PC to monitor the analog outpu...

Страница 8: ...l can be connected to the positive or negative of the power supply The maximum power output of each relay is 90W 5A Output control interfaces and load connection diagram Output control interfaces and...

Страница 9: ...ce value will be displayed Display Details The 9825 utilizes a 128 x 32 dot OLED display with adjustable LED back light The table below summarizes the display annunciators Basic Force Display Force Di...

Страница 10: ...ay mode PK Val Function Cycle among real time peak and valley display modes When in the setup menu Function Enter sub menu Decrease a value when used as a directional key Reset When in display mode Re...

Страница 11: ...mooths signal by averaging samples over a given span 1 Integer values from 1 to 100 Units Cal Basic Unit Select engineering units displayed lbf lbf ozf N kN g kg UNIT_NONE Nominal Capacity Sets displa...

Страница 12: ...8 bit None 8 bit Even 7 bit Even 7 bit Odd Port Mode Set port mode Demand Demand Continuous Protocol Set Protocol see protocol descriptions in the appendix Condec Condec ASCII Analog Out Output Type S...

Страница 13: ...he screen Press the Pk Val button to initiate the calibration process 4 The term Fixture will show on the screen At this point the load cell should be set into its fixturing but with no additional cal...

Страница 14: ...follow the steps below 1 Press and hold the Menu button to enter the setup menu A beep will sound as the setup menu is activated 2 Using the Menu button scroll until Calibration is displayed on the s...

Страница 15: ...t of the 9825 should be connected to whatever instrument will be accepting and measuring the analog signal 1 In the Analog Output setup menu scroll to Fine Tune and press the Pk Val button to initiate...

Страница 16: ...s less than the value of Input Point 4 but greater than the value of Input Point 3 The symbol will show on the display The OUT 4 relay will close Otherwise the symbol will show on the display and the...

Страница 17: ...relay will open 4 When the load is less than the value of Input ExtraLow The symbol will show on the display The alarm will sound The OUT 4 relay will close The display will send a warning message Oth...

Страница 18: ...Format 1 Continuous Mode ASCII In this mode of communication the indicator transmits the data frame continuously The load value in the frame is expressed in ASCII Interface Inc www interfaceforce com...

Страница 19: ...device PC will output demand command through the serial ports when the scale is in the normal loading status The demand command format is shown below The serial output data format is as follows Interf...

Страница 20: ...ndec Demand Output Demand Commands P Print T Tare Z Zero G Gross N Net Condec Continuous Output Note The starting address 40001 of MODBUS is not suitable for SIEMENS soft Interface Inc www interfacefo...

Страница 21: ...C 10 to 90 non condensing at F 10 to 90 non condensing POWER Supply VDC 24 VDC with supplied 120V 60Hz AC DC adapter or 9 36 VDC external supply Power Consumption W 6 RMS 8 Peak MECHANICAL Dimensions...

Страница 22: ...freight insurance and proper packaging to prevent breakage in transit Interface warranty does not apply to defects resulting from action of the buyer such as mishandling improper interfacing operatio...