950P
Portable Air Cleaning System
950P
Portable Air Cleaning System
#10065
18
© Intellipure 2015
#10065
19
© Intellipure 2015
fig-6
fig-5
Maintenance
Maintenance
V-
BANK MAIN FILTER REPLACEMENT
Lock the (2) front wheels to secure the unit.
(fig-1)
Remove the Filter Access Door located in the back of the unit using the four butterfly latches located on the
sides of the unit. Lift the latch wings (4) and turn counter-clockwise to open.
(fig-2)
Filter Access Door removal may require a firm pulling effort.
Using the Magnetic Bit Holder provided, alternate loosening the left and right Scissor Locking Mechanisms
located near the motor plate in the filter. Make sure the Scissor Locking Mechanisms are fully opened.
(fig-5)
Disconnect the High Voltage Lead from the High Energy Grid.
(fig-6)
Using a firm pull, remove the V-Bank Main Filter and High Energy Grid Assembly out of the filter access
opening.
(fig-7)
Inspect and clean the inside of the unit.
Refer to page 19 for proper instructions.
Separate the High Energy Grid from the
V-Bank Main Filter by removing the
six screws (4 corners and 2 in the center)
in the V-Bank Main Filter.
(fig-8)
Replace the V-Bank Main Filter.
Secure the V-Bank Main Filter to the bottom
of the High Energy Grid by using the
six screws (4 corners and 2 in the center)
in the V-Bank Main Filter.
Slide the V-Bank Main Filter and High Energy
Grid Assembly back into the unit. Pay
attention to the Flow Direction Arrow on
the Prefilter, the arrow should point
upwards towards the Supply/Outlet end
of the unit.
(fig-9)
Using the Magnetic Bit Holder provided,
alternate tightening the left and right
Scissor Locking Mechanisms located
near the motor plate in the filter.
CLEANING THE UNIT
Depending on contamination level, this procedure should be conducted about once a year or during each
filter change or if an electrical problem occurs.
Remove the V-Bank Main Filter and High Energy Grid Assembly
Refer to page 18 for proper instructions.
To clean the wires, use a fiber free swab, either dry or moistened with distilled or deionized water, clean
along the length of each wire, spring clip, and ceramic insulator.
Clean the inner surfaces of the High Energy Grid material using a lint cloth moistened with distilled or
deionized water.
Clean the High Energy Grid surface by vacuuming between the wires using a small vacuum attachment or
by using a lint free cloth, either dry or moistened with distilled or deionized water. Take care to avoid leaving
large fibers snagged on the wires, control grid, or other components of the High Energy Grid.
Make sure that any contaminant that falls to the bottom of the filter unit is removed.
In extreme cases, it may be necessary to remove the wires from the springs attached at each end to the power
distribution bars, thus allowing complete access to the inside of the High Energy Grid assembly for cleaning
as previously described. It is recommended that the manufacturer be contacted for detailed instructions
should this step appear necessary.
Vacuum the bottom of the filter seal plate section and of the filter unit to remove any particles dislodged during
cleaning. Reinstall the High Energy Grid, V-Bank Main Filter and Filter Access Door as previously instructed.
Reconnect the High Voltage Lead.
Ensure the Prefilter is firmly seated.
The unit will not run if the Prefilter is
not installed correctly.
Return the Filter Access Door. Lift the
latch wings (4) and turn clockwise
to close.
(fig-10)
fig-9
fig-10
Do not over-tighten the
Scissor Lock Clamps.
Do not use alcohol or cleaners that may damage acrylics, plastics or aluminum.
fig-7
fig-8