MB/TS25 PC211 UPGRADE Battery Charger
Operators Manual
7 BATTERY LIBRARY
The TS25 UPGRADE has 99 sets of built in data, which can be recalled quickly to perform
both charging and capacity testing. Recalling library data is covered in the sections on
capacity testing and charging.
7.1 RETRIEVING A LIBRARY ITEM
Press the
BATT
key. The display will show the current library item selected and
the rightmost digit will be flashing. Enter the library number by using the
and
keys and the
EXTD
key to move to the next digit. Only library numbers 01 to
99 can be selected.
The leftmost digit is used to select which library database the item is to be
retrieved from.
'0'
Indicates that the item is to be retrieved from the user data base where
items can be modified.
'1'
Indicates that the item is to be retrieved from the units internal fixed
battery library. Items from this database cannot be written back.
As supplied or when a library reset is performed the modifiable database is the
same as the un-modifiable database.
7.2 ENTERING OR MODIFYING A LIBRARY ENTRY
In order for a library item to be modified it must be recalled first, modified, and
then written back.
Recall the library item you wish to modify using the procedure above.
Follow the procedures for manual charging or capacity testing to set the
ASET
,
VSET
and
TSET
and
PROC
parameters.
Once these have been reviewed the battery library data can be written back by
pressing the
ENTR
key rapidly twice. This means the
ENTR
key must be pressed
twice within 0.6 Second. The display will briefly show
LIB
to confirm that the item
has been applied to the units’ internal memory.