15

1.  Maintain Moisture: 

Immediately after the surgical procedure, place the 

dermatome hand piece, width clip, guard, and screw fasteners in an instrument 
container and apply a transport foam or gel product (e.g., Steris PRE-Klenz®) 
specifically intended to keep soil moist on surgical instruments. Also introduce 
some of the foam or gel product into the oscillating pin recess of the cutting 
head. Transport the tray containing the soiled dermatome in an impervious 
plastic bag or container to the decontamination environment. Keep the outside 
of the container clean. 

2.  Enzymatic Soak:

 Prepare an enzymatic detergent solution (e.g., ASP® ENZOL®) as 

per the manufacturer’s recommendation. Remove the two hex nuts fastening the 
dermatome’s cutting head to the motor housing. Immerse only the dermatome’s 
cutting head, width clips, guard, and screw fasteners in the enzymatic solution. 

Do not immerse the dermatome’s motor housing.

 Immersion of the motor 

housing can cause permanent damage to its electrical components.

CAUTION:

 Removing the dermatome’s cutting head from the motor housing 

will result in exposing the oscillating pin and motor drive shaft. The contour 
of the oscillating pin is factory set to provide proper dermatome performance. 
Care must be taken in handling the motor housing in this state. Dropping the 
motor housing or hitting the oscillating pin can cause damage that will result in 
the instrument being out of calibration upon re-assembly.

3.  Rinse: 

After the enzymatic detergent manufacturer’s recommended soak period, 

remove dermatome’s cutting head, width clips, and screw fasteners from the 
detergent solution and thoroughly rinse the items with tap water.

4.  Clean Instruments:

 Prepare a cleaning solution appropriate for surgical 

instruments (e.g., ASP ENZOL) as per the manufacturer’s instructions. Using a 
small clean soft-bristled brush, remove soil from all surfaces of the dermatome’s 
cutting head, width clips, guard, and screw fasteners while they are submerged 
in the solution. Take particular care in removing soils from the dermatome’s 
cutting head features containing crevices and sharp inside corners. Clean the 
dermatome’s motor housing by wiping with a soft clean cloth moistened with 
the detergent solution. Pay special attention to removing soil from the area of 
the oscillating pin and drive shaft. Clean the power cord by wiping with a soft 
clean cloth moistened with the detergent solution.

5.  Rinse: 

Thoroughly rinse all dermatome components under gently running tap 

water and wipe with a dry, soft, clean cloth. Do not immerse the dermatome 
components.

Cleaning And Decontamination (Continued)

Содержание Padgett SB

Страница 1: ...Integra Padgett Model SB Slimline Dermatome...

Страница 2: ...line 3 Classifications 4 Specifications 4 Warnings 5 Precautions 6 International Symbols 8 Calibration 9 The Power Supply 10 How to Use the Model SB Slimline Dermatome 11 Maintenance And Servicing 13...

Страница 3: ...rmatome is delivered to you in a sturdy carry case with the power supply power supply cord 13 ft 3 96 m hand piece cord guard 2 3 and 4 inch 5 08 7 62 and 10 16 cm width clips screwdriver calibration...

Страница 4: ...ply 2 Hand Piece Cord 3 Hand Piece 4 Blade Clips 5 Calibration Guide 6 Screwdriver 7 Power Supply Wall Cord 2 3 4 5 1 Catalog Description Optional Accessories 3539 701 Dermatome hex nuts 2 pcs 3539 26...

Страница 5: ...DO NOT use in the presence of flammable anesthetics Specifications Maximum Graft Width SB 4 10 16 cm Input Voltage 100 250Vac Input Frequency 50 60Hz Output Volts 48Vac Output Current 1A Operating Tem...

Страница 6: ...zation and subsequent reuse will dull the blades and may result in cross contamination or impaired function of the product Sharpness of the blades is not guaranteed with repeated use Any blade once us...

Страница 7: ...Refer servicing to an authorized Integra Padgett Dermatome Repair Center Use Model SB hand piece with Model SB power supply and hand piece cord ONLY Prior to use visually inspect the dermatome hand p...

Страница 8: ...nnectors are to fit properly with their mating receptacles 4 Connect power supply power cord to wall receptacle and turn power supply switch on The indicator light on the power supply is to illuminate...

Страница 9: ...ol represents the international symbol for cutting This symbol represents the international symbol for Caution consult accompanying documents This symbol represents the international symbol for Altern...

Страница 10: ...n lies anywhere within the zone clearly marked on the calibration guide the dermatome is properly calibrated If the pin lies on the edge or outside the zone the dermatome should not be used and should...

Страница 11: ...light indicating that the dermatome is ready for use whenever the light is on Please note there may be up to a two second delay when the power switch is first activated before the green light illumina...

Страница 12: ...grasp the silver cap CAUTION Blades are single use ONLY DO NOT reuse blades The on off switch is on top of the hand piece and is operated with the thumb Push the switch forward for continuous mode Pu...

Страница 13: ...of the motor should be carried by the hand that holds the dermatome to maintain a reasonably light pressure between the cutting edge and the patient s body The instrument should be held at a 25 45 ang...

Страница 14: ...Service prior to returning product The product is required to be properly cleaned and sterilized before it is returned to the repair center with documentation verifying sterilization Failure to steri...

Страница 15: ...ting head and the edge of the guide bar Only remove the components as instructed below Further disassembly of the instrument is likely to cause permanent damage The dermatome contains no user servicea...

Страница 16: ...r dermatome performance Care must be taken in handling the motor housing in this state Dropping the motor housing or hitting the oscillating pin can cause damage that will result in the instrument bei...

Страница 17: ...les in the motor housing flange with the threaded studs of the cutting head Note that the thumbswitch on the motor housing is to be positioned on the top of the unit Tip the motor housing to create su...

Страница 18: ...re This is especially important when utilizing a flash sterilization cycle 100 Ethylene Oxide Sterilization Precondition the chamber to the specified exposure temperature and 55 relative humidity for...

Страница 19: ...ENT SHALL INTEGRA BE LIABLE FOR ANY INCIDENTAL INDIRECT CONSEQUENTIAL OR PUNITIVE DAMAGES IN CONNECTION WITH THE ACQUISITION OR USE OF ANY INTEGRA PRODUCT Further this warranty shall not apply to and...

Страница 20: ...ses Inc Pre Klenz is a trademark of Steris Corporation ASP Enzol and Sterrad are trademarks of Johnson and Johnson 2020 Integra LifeSciences Corporation All rights reserved Printed in USA PL 00196 Rev...

Отзывы: