8
EN – ENGLISH
4. Follow the instructions listed below and set washer
machine to these exact parameters:
Phase
Time
(Min.)
Water Tem-
perature
Detergent and
Concentration
Pre-wash 1
4:00
Cold water
N/A
Enzyme
Wash
4:00
Hot water
Neutral pH
enzymatic (e.g.
Endozime® AW
Triple Plus with
APA, Ratio
1:128)
Phase
Time
(Min.)
Water Tem-
perature
Detergent and
Concentration
Wash 1
10:00
60.0°C
(140° F)
Neutral pH
detergent
(e.g. Renu-Klenz™ ,
Steris, Ratio 1:256)
Rinse 1
0:30
Hot water
N/A
Thermal
Rinse**
2:00
82.2°C
(180° F)
N/A
** Optional phase for disinfection of components –
minimum water temperature as indicated or per worker
manufacturer specifications for the thermal rinse cycle.
NOTE
– Any deviation from this guideline could result in
damage to the equipment as well as improper cleaning
results.
Rinse with purified water. Do not perform if parameters
cannot be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible
organic debris or residue from cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is expected
to be as long as 7 years.
Maintenance and Care
To ensure proper function and to extend the life and
performance of the equipment, Integra LifeSciences
recommends the following:
In the absence of proper care and servicing of the device,
negative effects may be seen after repeated processing
over time which may lead to reduced performance.
Contact information: See the Service and Repair section
for contact information on how to return your device for
periodic servicing and to request periodic inspections.
See Inspection and/or Service notes section for routine
checks to be performed on the device.
NOTE
– Any serious incident that has occurred in relation
to the device for the user and/or the patient should
be reported to the manufacturer and the competent
authority of the member state in which the user and/or
patient is established.
Device Disposal
NOTE: Follow hospital procedures for disposal of this device.
Integra LifeSciences Warranty Statement
INTEGRA LIFESCIENCES CORPORATION (“INTEGRA”)
warrants to the original purchaser only that each new
MAYFIELD product is free from manufacturing defects in
material and workmanship under normal use and service
for a period of one year (except as otherwise expressly
provided as to accessory items) from the date of delivery
by INTEGRA to the first purchaser, but in no event beyond
the expiration date stated on any product labeling.
• Surgical instruments are guaranteed to be free
from defects in material and workmanship when
maintained and cleaned properly and used normally
for their intended purpose.
• Any covered product that is placed by INTEGRA
under a lease, rental or installment purchase
agreement and that requires repair service during
the term of such placement agreement shall be
repaired in accordance with the terms of such
agreement.
Recommended Action
Recommended
Frequency
Return the device to the Integra
LifeSciences Repairs department for
detailed inspection and servicing.
Once / year
Request that Integra NeuroSpecialists
perform routine inspections of the
device
Twice / year
Содержание MAYFIELD A1015
Страница 2: ...2 This page is intentionally left blank...
Страница 3: ...3 MAYFIELD Crossbar Adaptor A1015 A1016 Instruction Manual EN English...
Страница 11: ...11 Barre de Croisement MAYFIELD A1015 A1016 Mode d emploi FR Fran ais...
Страница 19: ...19 Barra trasversale per unit di base MAYFIELD A1015 A1016 Manuale di istruzioni IT Italiano...
Страница 27: ...27 MAYFIELD Stangen Adapter A1015 A1016 Gebrauchsanleitung DE Deutsch...
Страница 35: ...35 Adaptador de barra cruzada MAYFIELD A1015 A1016 Manual de instrucciones ES Espa ol...
Страница 43: ...43 43 MAYFIELD dwarsstangadapter A1015 A1016 Gebruikershandleiding NL Nederlands...
Страница 51: ...NL NEDERLANDS 51 Deze pagina is bewust leeg gelaten...