Powering the ICP Module & 

Calibrating to the Patient Monitor

Patient monitor will read >250 mmHg

until ICP Sensor is zeroed and active

Connect DirectLink® ICP Module

 to the  

bedside Patient Monitor (PM). Panel lights 
will flash in sequence during self-test. 

Wait

 for SOLID AMBER status light before 

continuing.

Press

 ARROW BUTTON until the 0 mmHg 

reference signal indicator illuminates. 

Zero

 patient monitor per manufacturer’s 

instructions.

Press

 ARROW BUTTON until the 100 mmHg 

reference signal indicator is illuminated. 

Check

 that the value displayed on the patient 

bedside monitor is 100.

Exit

 calibration mode by pressing ARROW 

BUTTON until neither green reference signal 
indicator is lit and status light is SOLID 
AMBER.

2

PATIENT MONITOR SETUP

Содержание Codman DirectLink

Страница 1: ...Codman Microsensor ICP Transducer to a Patient Monitoring System Reference Signal Indicators Status Lights Arrow Button Patient Monitor Calibration Zero Button ICP Sensor Zeroing Output Connection To...

Страница 2: ...AMBER status light before continuing Press ARROW BUTTON until the 0 mmHg reference signal indicator illuminates Zero patient monitor per manufacturer s instructions PressARROWBUTTONuntilthe100mmHg ref...

Страница 3: ...ID AMBER prior to continuing Prepare the Codman Microsensor ICP Transducer for zeroing per the user manual Press and Hold the ZERO BUTTON for approximately 8 seconds Zeroing begins when amber status l...

Страница 4: ...if needed Press ARROW BUTTON and ZERO BUTTON simultaneously ACKNOWLEDGE FUNCTION to confirm sensor zero value and continue monitoring SOLID GREEN status light will illuminate after 4 seconds and ICP...

Страница 5: ...T to try again If problem persists patient monitor may be incompatible or DirectLink ICP Module needs to be replaced DURING USE A damaged cable or sensor may have been detected 1 Disconnect all cables...

Страница 6: ...k Cables are positioned away from other electrical equipment hold sensor steady and repeat zeroing process If problem persists it is likely a bad ICP Extension Cable ICP Sensor or electrical interfere...

Страница 7: ...r notice to modify the products in order to improve their quality n Warning Applicable laws restrict these products to sale by or on the order of a physician n Consult product labels and inserts for a...

Отзывы: