background image

 

Rotate the battery enclosure in an anti-clockwise (counter clockwise) direction to 

pre-twist the wires before screwing on (see fig 7). 

 

Ensure that no wires are going to be captured or damaged by screwing on the  

battery enclosure. 

 

Bring the thread of the battery enclosure into contact with the thread of the  

logger. 

 

Make sure the battery enclosure and the logger body are exactly aligned. 

 

With small positive pressure, screw the battery enclosure into the logger body (see 

fig 8). 

 

If the battery enclosure rotates more than one revolution without the threads  

engaging, stop and repeat the process. 

 

Check that the O ring on the communication socket for damage, and replace if  

necessary (see fig 9). 

 

Connect the logger to a computer and check the battery voltage on the real-time 

screen. 

 

If everything is OK, replace the label with the new label provided (see fig 10). 

Fig 1: 

Unscrew battery enclosure  

Anti-clockwise. 

Fig 2: 

Identify the Battery and  

Communications wires. 

Note: 

If any of the wires are red, black,  

orange and yellow the battery enclosure will 

need to be replaced with the Stainless Mk4 

tube (battery enclosure).

 