11. TROUBLESHOOTING (2)

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD shows low 

battery symbol 

Low Battery

Change the 

batteries

LCD shows “Er 0”

Pressure system 

is unstable before 

measurement

Don’t move and 

try again.

LCD shows “Er 1”

Fail to detect systolic 

pressure

LCD shows “Er 2”

Fail to detect diastolic 

pressure

LCD shows “Er 3”

Pneumatic system 

blocked or cuff is too 

tight during inflation

Apply the cuff 

correctly and try 

again

LCD shows “Er 4”

Pneumatic system 

leakage or cuff is too 

loose during inflation

LCD shows “Er 5”

Cuff pressure above 

300mmHg

Measure again 

after 

five

 minutes. 

If the monitor is 

still abnormal, 

please contact the 

local distributor or 

the factory.

LCD shows “Er 6”

More than 3 minutes 

with cuff pressure 

above 15 mmHg

LCD shows “Er 7”

EEPROM accessing 

error

LCD shows “Er 8”

Device parameter 

checking error

LCD shows “Er A” Pressure sensor 

parameter error

No response when 

you press button 

or load battery.

Incorrect operation or 

strong electromagnetic 

interference.

Take out batteries 

for five minutes, 

and then reinstall 

all batteries.

Содержание IMC-734

Страница 1: ...Owner s manual Blood Pressure Monitoring System IMC 734...

Страница 2: ......

Страница 3: ...NECTING THE CUFF TO THE MONITOR 9 4 APPLYING THE CUFF 9 5 BODY POSTURE DURING MEASUREMENT 10 6 TAKING YOUR BLOOD PRESSURE READING 10 7 DISPLAYING STORED RESULTS 11 8 DELETING MEASUREMENTS FROM THE MEM...

Страница 4: ...nually day and night The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the day...

Страница 5: ...Bar START button Date Time Display Alternating Blood Pressure level Classification Indicator Diastolic Pressure Pulse Rate Display Alternating Ready to inflate indicator Irregular Heartbeat Symbol Lo...

Страница 6: ...sensor blood pressure and pulse rate can be measured automatically and non invasively The most recent 60 measurements can be stored in the memory with date and time stamp The Electronic Sphygmomanomet...

Страница 7: ...uring accuracy 3mmHg 13 Heart pulse rate range 40 180times min accuracy 5 14 Environmental temperature for operation 5 C 40 C 15 Environmental humidity for operation 90 16 Environmental temperature fo...

Страница 8: ...ung children Consult your physician or other health care professionals before use on older children 8 Blood pressure measurements determined by this monitor are equivalent to those obtained by a train...

Страница 9: ...egulations at the end of their usage There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of pulse period 25 2 The difference of adjacent pulse peri...

Страница 10: ...at first See picture 2 2 Press the button START repeatedly the day hour and minute will blink in turn See picture 2 3 picture 2 4 picture 2 5 While the number is blinking press the button MEM to incr...

Страница 11: ...of your body on a desk or table with the palm up If the cuff is correctly placed you can read the LCD display c The cuff must be neither too tight nor too loose Note Please refer to the cuff circumfe...

Страница 12: ...ow resting on a chair or table c The cuff should be at the same level as your heart d Placing your arm on the monitor s plastic case as shown in the picture to aid in maintaining a correct and still a...

Страница 13: ...measurement the monitor will turn off automatically after 1 minute of no operation Alternatively you can press the START button to turn off the monitor manually e During measurement you can press the...

Страница 14: ...the button START to turn off the monitor manually 8 DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying keep on pressing button MEM for three seconds all results in the memory bank wi...

Страница 15: ...t intended to provide a basis of any type of rush toward emergency conditions diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels o...

Страница 16: ...LCD shows Er 4 Pneumatic system leakage or cuff is too loose during inflation LCD shows Er 5 Cuff pressure above 300mmHg Measure again after five minutes If the monitor is still abnormal please conta...

Страница 17: ...mance should be checked every 2 years or after repair Please contact the service center 7 If the monitor becomes dirty please clear it with a soft dry cloth Do not use any abrasive or volatile cleaner...

Страница 18: ...COMPILES WITH MDD93 42 EEC REQUIREMENTS Symbol for DATE OF MANUFACTURE Symbol for EUROPEAN REPRESENTATION WARRANTY INFORMATION Only charge the cost of components and transport SERVICE CENTER Inspire M...

Страница 19: ...environment guidance RF emissions CISPR 11 Group 1 The IMC 734 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nea...

Страница 20: ...l Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with s...

Страница 21: ...ment should be used no closer to any part of the IMC 734 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommend...

Страница 22: ...e RF compliance level above the IMC 734 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the IM...

Страница 23: ...ot listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the...

Страница 24: ......

Отзывы: