Inspire IMC-734 Скачать руководство пользователя страница 11

8

9

3. CONNECTING THE CUFF TO THE MONITOR

The cuff is attached to the monitor when it is packaged. Should the 

cuff become unattached, align the two plugs and four brackets of 

the cuff with the plug sockets and bracket sockets of the monitor 

and press the cuff to the monitor until the plugs and brackets are 

securely attached. 

4. APPLYING THE CUFF

a. Place the cuff around a bare wrist 1-2cm above the 

wrist joint on the palm side of the wrist. 

b. While seated, place the arm with the cuffed wrist in 

front of your body on a desk or table with the palm up. 

If the cuff is correctly placed, you can read the LCD 

display.

c. The cuff must be neither too tight nor too loose. 

Note: 

• Please refer to the cuff circumference range in 

“SPECIFICATIONS” to make sure that the appropriate 

cuff is used.

• Measuring on same wrist each time.

• Do not move your arm, body, or the monitor during measurement.

•  Stay  quiet,  calm  for  5  minutes  before  blood  pressure 

measurement.

• Please keep the cuff clean. If the cuff becomes dirty, remove 

it from the monitor and clear it by hand in a mild detergent, then 

rinse it thoroughly in cold water. Never dry the cuff in clothes dryer 

or iron it. Clean the cuff after the usage of every 200 times is 

Содержание IMC-734

Страница 1: ...Owner s manual Blood Pressure Monitoring System IMC 734...

Страница 2: ......

Страница 3: ...NECTING THE CUFF TO THE MONITOR 9 4 APPLYING THE CUFF 9 5 BODY POSTURE DURING MEASUREMENT 10 6 TAKING YOUR BLOOD PRESSURE READING 10 7 DISPLAYING STORED RESULTS 11 8 DELETING MEASUREMENTS FROM THE MEM...

Страница 4: ...nually day and night The highest value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the day...

Страница 5: ...Bar START button Date Time Display Alternating Blood Pressure level Classification Indicator Diastolic Pressure Pulse Rate Display Alternating Ready to inflate indicator Irregular Heartbeat Symbol Lo...

Страница 6: ...sensor blood pressure and pulse rate can be measured automatically and non invasively The most recent 60 measurements can be stored in the memory with date and time stamp The Electronic Sphygmomanomet...

Страница 7: ...uring accuracy 3mmHg 13 Heart pulse rate range 40 180times min accuracy 5 14 Environmental temperature for operation 5 C 40 C 15 Environmental humidity for operation 90 16 Environmental temperature fo...

Страница 8: ...ung children Consult your physician or other health care professionals before use on older children 8 Blood pressure measurements determined by this monitor are equivalent to those obtained by a train...

Страница 9: ...egulations at the end of their usage There are 2 conditions under which the signal of IHB will be displayed 1 The coefficient of variation CV of pulse period 25 2 The difference of adjacent pulse peri...

Страница 10: ...at first See picture 2 2 Press the button START repeatedly the day hour and minute will blink in turn See picture 2 3 picture 2 4 picture 2 5 While the number is blinking press the button MEM to incr...

Страница 11: ...of your body on a desk or table with the palm up If the cuff is correctly placed you can read the LCD display c The cuff must be neither too tight nor too loose Note Please refer to the cuff circumfe...

Страница 12: ...ow resting on a chair or table c The cuff should be at the same level as your heart d Placing your arm on the monitor s plastic case as shown in the picture to aid in maintaining a correct and still a...

Страница 13: ...measurement the monitor will turn off automatically after 1 minute of no operation Alternatively you can press the START button to turn off the monitor manually e During measurement you can press the...

Страница 14: ...the button START to turn off the monitor manually 8 DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying keep on pressing button MEM for three seconds all results in the memory bank wi...

Страница 15: ...t intended to provide a basis of any type of rush toward emergency conditions diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels o...

Страница 16: ...LCD shows Er 4 Pneumatic system leakage or cuff is too loose during inflation LCD shows Er 5 Cuff pressure above 300mmHg Measure again after five minutes If the monitor is still abnormal please conta...

Страница 17: ...mance should be checked every 2 years or after repair Please contact the service center 7 If the monitor becomes dirty please clear it with a soft dry cloth Do not use any abrasive or volatile cleaner...

Страница 18: ...COMPILES WITH MDD93 42 EEC REQUIREMENTS Symbol for DATE OF MANUFACTURE Symbol for EUROPEAN REPRESENTATION WARRANTY INFORMATION Only charge the cost of components and transport SERVICE CENTER Inspire M...

Страница 19: ...environment guidance RF emissions CISPR 11 Group 1 The IMC 734 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nea...

Страница 20: ...l Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with s...

Страница 21: ...ment should be used no closer to any part of the IMC 734 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommend...

Страница 22: ...e RF compliance level above the IMC 734 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the IM...

Страница 23: ...ot listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the...

Страница 24: ......

Отзывы: