9

10

Shutdown

 

Time: To  maximize  the  life  of  the  battery,  power  is 

off  after  a  while  of  non-use.  The  time  can  be  set  under  this 
menu.  The  range  is  01-99  minutes.  When  set  to  “99”  the 
system  will  always  on.  Touch  “UP”  to  adjust  the  value,  touch 
“DOWN”  to  shift  to  the  next  position,  touch  “EXIT”  to  abort, 
and touch “ENTER” to confirm and back.

Backlight  Time:  Under  this  menu,  the  continued  lighting  time 
of  the  backlight  can  be  set.  Touch UP”  or  “DOWN”  keys  to 
shift  to  the  next  position,  touch  “EXIT”  to  abort,  and  touch

ENTER” to confirm and back.

  “

 

Key

 

Sound:  The  key  sound  can  be  turn  on  or  off  by  this 

setting.  Touch  “UP”  or  “DOWN”  keys  to  shift  to  the  next 
position, touch “EXIT” to abort, and touch “ENTER” to confirm 
and back.
Date/Time: If necessary, the system time can be set under this 
menu. Touch  “UP”  to  adjust  the  value,  touch  “DOWN”  to  shift 
to the next position, touch “EXIT” to abort, and touch ENTER” 
to confirm and back.

 “

Contrast:  Due  to  the  use  of  different  ambient  temperatures 
and  light  conditions,  it  requires  to  adjust  the  contrast  of  LCD 
in order to making for observing and reading data. Touch “UP” 
or  “DOWN”  keys  to  adjust  the  value,  touch  “EXIT”  to  abort, 
and touch “ENTER” to confirm and back.
Key  Settings:  Under  this  menu,  It  is  only  to  set  the  default 
function of the key “SEND/EXIT” under the measure interface. 
Print  the  recent  data  or  stores  the  current  display  value  can 
be selected.

Default

 

Settings:  When  the  artificial  error,  and  do  not  know 

how to restore, the gauge can be restored to factory settings. 
Carefully use this function!

Calibration: If necessary, user can self calibrating the gage under 
this  menu.  There  are  two  ways  of  calibration,  With  Five  Point  or 
Modulus Modify.

5

With  Five  Point:  A  prompt  window  will  be  pushed  out  under 
This menu. Calibration will be executed by selection of “YES”, 
and  this  operation  will  be  aborted  by  selection  of  “NO”  or 
touch “EXIT”. 20%, 40%,60%,80%,100% of the rated capacity
should be recommended. All of the five points can be same.

Содержание ISF-DF Series

Страница 1: ...ISF DF SERIES www insize com OPERATION MANUAL DIGITAL FORCE GAGE ISF DF SERIES E V0...

Страница 2: ...the range of stated temperature and humidity or it may result in apparatus failure Operating Temperature 10 C 40 C Relative Humidity 20 80 Storage Temperature 20 C 50 C Relative Humidity 5 90 1 Measur...

Страница 3: ...e value The Force Gage has multi level menu interface 1 The Force Gage has multi level menu interface Under the measure interface touch MENU can enter the main menu interface Testing Setup This menu i...

Страница 4: ...confirm and back Printing Setup This menu item contains three selectable items Print Recent Print Selected and Print All The data saved in memory can be output to printer through the serial port RS23...

Страница 5: ...later data unit Browse Z to A The last page data will be showed when enter this menu other like Browse A to Z the same Browse Selected Under this menu the user can choose the data to browse The range...

Страница 6: ...TER to confirm and back Contrast Due to the use of different ambient temperatures and light conditions it requires to adjust the contrast of LCD in order to making for observing and reading data Touch...

Страница 7: ...ware and the serial number 6 Charging The Digital Force Gage is supplied with a set of 3 Nickel Metal Hydride AAA rechargeable batteries which are supplied fully charged to allow use straight in the i...

Отзывы: