OPERATION
Note! – In “Normal Operation” the hinge
cover is to remain tightly screwed closed. Under no
circumstance is it necessary for the operator to open
the enclosure during normal operation.
RUN Mode
The RUN mode is the normal operating mode of the analyzer and is entered upon power-up. When
the Run mode is entered the analyzer will determine what types of sensors are attached to each channel.
The display is continuously updated with the current measurement values. Also, the analog output and
the relays are updated according to the current conditions and their programmed functions. In the event
of an error or alarm condition the display will indicate the problem in plain English text.
The supplied sensors have been calibrated at the factory. Sensor information for the Model 15/15L
TSS sensors are stored in non-volatile memory of the analyzer. If a Model 15/15L is connected to the
analyzer and the sensor information does not match the information stored in the analyzer, the analyzer
will display a configuration message. The configuration message for the Model 15/15L TSS sensor is
“**Zero sensor**”. These messages will appear whenever the sensor is changed.
While in the RUN Mode, the time to next scheduled clean cycle can be viewed by pressing and
holding either arrow key. A clean cycle can be demanded by pressing the ENTER key while in the RUN
Mode, see the section on Demand Clean.
MAIN MENU
The MAIN MENU is accessed by pressing the “MENU” key while in the RUN mode of operation.
There are three options available from the main menu. Use the arrow keys to switch between RUN,
SETUP, and TEST, and then press the "ENTER" key to select.
SETUP Mode
This mode of operation allows the user to customize the unit to the specific operation and needs of
the facility. There are a total of five subcategories that may be adjusted.
Operation of the SETUP MODE proceeds as follows:
From the MAIN MENU, select SETUP and the five menu options will be displayed. Use the
“ARROW” keys to move the cursor to the desired SETUP option, and then press the “ENTER” key. A
menu with five options will be displayed. The options are:
RELAYS
ANALOG OUTPUT
MODBUS
CHANNEL ENABLE
SENSOR
Содержание MPA-48
Страница 22: ...2 high order byte...
Страница 30: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ...Suite 107 Slidell LA 70458 InsiteIG 80 Whisperwood Blvd...
Страница 45: ...1550 W Lindburg Dr Slidell LA 70458 Insite IG...
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......