4.4 NIBP Error and Possible Cause of Error
Error
Cause
SysErr
Self-test fail
SysErr2
NIBP module system error
CuffLoose
Cuff is too loose, or cuff not connected
CuffErr
Use small animal cuff under vet mode
Leakage
Valve or gas circuit leak
PressErr
NIBP Valves are not working appropriately
Weak
Patient’s pulse is too weak or cuff is loose