40
should be placed at the same level as the
patient’s heart.
You can adjust the
measurement results as below:
◆
If the cuff is placed higher than the heart level, add
0.75mmHg (0.10kPa) per each centimetre gap.
◆
If it is placed lower than the heart level, deduct
0.75mmHg (0.10kPa) per each centimetre gap.
4. Check the correct NIBP measurement mode that is
suitable for your species type.
5. Press the Function 2 button to start testing
4.3 Operation Instruction
1.
Perform Automatic Measurement
User can set the measurement interval time to start
Содержание Vet Vital Signs
Страница 1: ...Monitor Operator s Manual ...
Страница 2: ......
Страница 12: ...12 Chapter II Main Parts Accessories 2 1 Button and Indicator Light Fig 2 1 1 Button and Indicator Light ...
Страница 16: ...16 2 4 Ports Fig 2 4 1 Ports ...
Страница 17: ...17 2 5 Mounting Hole Fig 2 5 1 Mounting Hole ...
Страница 19: ...19 Chapter III Interface 3 1 Main Interface Note ...
Страница 23: ...23 Fig 3 6 System Menu ...
Страница 29: ...29 Choose OK System will return to Review Menu Fig 3 6 6 Review Menu ...
Страница 34: ...34 Fig 3 6 6 2 Temperature Trend Chart ...
Страница 37: ...37 cause Ischaemia Fig 4 2 1 Cuff Usage ...
Страница 47: ...47 Fig 4 5 1 Take off the rubber hose from the cuff ...
Страница 52: ...52 no Sensor Sp02 sensor not connected Sensor off No paw in sensor Searching Searching pulse ...
Страница 60: ...59 Small Animal 3digits 60 ...
Страница 63: ...63 Battery Lithium Ion Rechargeable Battery 3 6V 4 2Ah Work Time 8 hours Charge Time 6 hours ...
Страница 65: ...64 3 Select Next 4 Select Install 65 ...
Страница 66: ...64 5 Select Next 6 Select Finish 7 Press Close 66 ...
Страница 67: ...64 8 The below icon will appear on your desktop 67 ...
Страница 69: ...64 ...