Ventway Sparrow
User Manual
Page 39 of 144
8.6.2. INDICATORS
_
T
_
Patient trigger detected
_
C
_
Patient cough detected
Note:
If a cough is detected, the ventilator will
ignore patient triggers for one second.
_
Z
_
Pressure sensors are currently being zeroed
Note:
While zeroing sensors, the ventilator will
not detect patient triggers.
The bottom left corner of the display shows the active alarms indicator.
8.6.3. ALARMS
Caution:
It is the user’s responsibility to ensure that the
alarm
s are working
for instructions on testing
the alarms.
Normal operation
Alarm is silenced.
Note:
While the alarm is silenced (maximum 120
seconds) this indicator is displayed, and the num-
ber of active alarms is displayed in parentheses
next to the icon.
Note:
The icon serves as an activation button.
Pressing it will open a list of up to five active
alarms, ordered by priority.