19P03-DOC01E REV-3
30
Integrated scale (option)
To use the integrated scale, please find below the different functions added to the
interface of the nurse control in the foot of the bed.
Symbols
Button features.
[Enable] button:
To activate and navigate the panel and all its functions, push the enable
button and a selected function button simultaneously. This is a safety feature
to avoid unintended usage.
[Accuracy] button:
By default, the system is set to 500g accuracy, but by activating this button
the accuracy changes to 100g.
The approval covers 500g as default. 100g only serves as a guidance (not
approved according to EN45501). The LED above the button will light
when 100g accuracy is chosen by pressing the button.
The 0.1 status automatically times out and changes to 500g default
indication after 5 sec. (the time out setting is required by the test house).
Toggling the accuracy to 0.1 (100g) guiding measure can only take place
when a stable load is present (when the display is no longer flashing).
[Scale] button:
To measure the weight, push the enable button and the scale button
simultaneously. The maximum measurement area is 0 - 460 kg.
The scale system calculates the maximum weight according to the following
formula:
Weight max = SWL - Auto Compensation - Zeroing = Actual weight of the
patient on the display.
• SWL = max 460 kg
• Auto compensation range = 0 - 100 kg
• Zeroing range = 0 - 50 kg
Example of display indication:
Max. patient weight = 460 kg - 100 - 40 = 320 kg.
I.e. Max. 320 kg can be measured
The LED above the button will light-up when the button is activated.
Never exceed the loading capacity of the bed even if the scale has a higher
capacity.