5.5 Probe Tip Cover
The meter ships with a re-usable probe tip cap. This cap can be
stored in the meter’s handle while the meter is in use. The use
of this probe tip cap is recommended for protecting the probe tip,
personal, and vehicles during transport.
5.6 Probe Tip Replacement
The probe tip is replaceable. It can be unscrewed from the shaft
by using pliers to grip the shaft and cone respectively. The cone
is attached with removable type thread locker, so some force
may be required to loosen it. The probe tip should be replaced
when the diameter of the cone’s base measures less than
0.490in.(12.45mm) diameter.
6.0
Troubleshooting & Maintenance
Errors:
If an error message (
Err2
or
Err3
) is displayed during the zero
process, then retry the zero button and assure the meter’s shaft and tip
are hanging vertically below the meter and not touching anything. If the
error message continues to show after repeated zeroing attempts, then
the meter must be serviced.
Cleaning:
Clean meter head and probe shaft with a damp cloth when
required.