NOTE: The EUT is the a.c. mains voltage prior to application of the test
level.
The following phenomenon is still fulfill the requirement of basic safety
and essential performance.
*UT:230V ~/50H z.The pressure of the EUT is deviation the normal value
but the value is still more than 10psi when flow is 4.5l/min.
**UT:230V ~/50Hz.The EUT stop working when adding 0%UT,but the
EUT can restore its normal mode automatically.
Use of this equipment adjacent to or stacked with other equipment
should be avioded because it could result in improper operation. If such
use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.
Portable RF communications equipment(including peripherals such as
antenna cables and external antennas) should be used no closer than
30cm(12 inches to any part of this devie, inluding cables specificed by
the manufacturer.
Otherwise, degradation of the performance of this equipment could
result. Under the test condition specified in immunity, the product can
provide the basic safety and essential performance.
If the essential performance is lost or degraded,additional measures are
necessary,such as reorienting or relocating the device.