DISplay:FORMat
[
0..5
]
Query or set the number of numeric fields on the panel.
Print displayed values.
FORMat:START <N>
Query or set the range for data array transfer
:END <N>
Range of N for harmonic values is 1 to 63
Range of N for analog inputs is 0-7.
If the value specified is out of range or start>end the
correction is done when values are queried using
VOLT:FFT?, CURR:FFT?, POW:FFT?, IMP:FFT?,
IMP:MAG, IMP:ANG, or AINP?.
AINPort
<
F
>
Set display field for analog input port previously selected
by the FORMat:START/END command.
AINPort?
Query all analog input port values identified by the
FORMat:START/END command. Range of allowed ports
is 0 to 7.
VERsion?
Query form only. Returns software version
LOCk
Locks the instrument’s front panel controls. The query
form returns YES or NO whether the controls are locked or
not.
UNLock
Unlock the instrument front panel controls.
RS232?
Query form only, returns all settings. Output format is
BAUD;PARITY;TERM;HAND
RS232:BAUD
2400
Query or set baud rate
4800
9600
19200
38400
57600
115k2
6-11
6-12
:PARITY
None
Query or set parity mode
Even
Odd
:
TERMinator
CR
Query or set command terminating characters.
LF
CRLf
:HANDshakes
None
Query or set handshake mode
Xon
GPIB:ADDRess
[
0..30
]
Query or set GPIB address
*ESE
[
0..255
]
Query or set the Event Status Enable register
*SRE
[
0...255
]
Query or set the Service Request Enable register
*STB?
Query the STatus Byte register (IEEE-488 only)
*ESR?
Query the Event Status Register (IEEE-488 only)
*RST
Resets instrument
*OPC
Set ISR bit 0 if no printing AND no measurement are pending
(GPIB
only).
*OPC?
Returns 1 if no printing AND no measurement are pending.
*TST?
Performs self test, returns zero if successful
*WAI
Suspends command execution until previous commands are complete.
*TRG
Forces a running measurement to become pending if in RUN mode.
Forces a measurement if in STOP mode.
*IDN?
Returns identification string in form:
<
Vendor, Model, Serial No,
Firmware version
>
.
*CLS
CLear status (no query form), (IEEE-488 only)
ERRor?
Query the last error code
ERROR CODES DEFINITIONS:
102 Syntax
Error
The command was not recognized. ESR bit 5 is set (CoMmand Error)
110
Command header error
A command followed by ‘?’ was sent were no query form is available. And conversly: no
‘?’ followed a query form only command. ESR bit 5 is set (CoMmand Error).
111
Header separator error
Attempted to descend the command hierarchy at a place where there wasn’t any
subcommand. ESR bit 5 is set (CoMmand Error).
Содержание SM201
Страница 28: ......