T-CAM 380 P-SERIES USER MANUAL
40
THE DEVICE WILL PERFORM A NUC CORRECTION EACH TIME A NEW
TEMPERATURE SCALE IS SELECTED.
A CAMERA RESTART IS REQUIRED AFTER CHANGING MODES.
IF MORE THAN 15% OF THE TEMPERATURES WITHIN THE FIELD OF VIEW
EXCEED 140 ℃, THE DEVICE WILL AUTOMATICALLY SWITCH TO THE
HIGH TEMPERATURE MEASUREMENT SCALE; IF MORE THAN 95% OF THE
TEMPERATURE IN THE FIELD OF VIEW ARE BELOW 120℃, THE DEVICE WILL
AUTOMATICALLY SWITCH TO THE STANDARD RANGE SCALE.
8-30
Temperature Scales
The default temperature scale of the device is -20 °C to 150 °C (-4 °F to 302 °F).
A high temperature calibration is also available. Change the temperature scale
as follows:
1.
Press the Menu button or tap the touchscreen to display the Main Menu.
2. Select the Settings icon:
3. Press the OK button to open Sub-menu.
4. Select Temperature Measurement Gear.
5. Press the OK button.
6. Select a temperature scale.
7. Press the OK button.
8. Press the Previous button continuously to exit the menu.
8-29-3
Live Streaming to a Computer
Users can synchronize the picture in the camera to the PC software for real-time
display and analysis. Stream directly to a computer as follows.
1.
Press the Menu button or tap the touchscreen to display the Main Menu.
2. Select the Settings icon:
3. Press the OK button to open Sub-menu.
4. Select USB Mode
5. Press the OK button.
6. Select USB Camera.
7. Press the OK button.
8. Turn the device Off, then On again. Image feed should now be viewable on
the external computer screen.