Страница 2: ...tings QC Test Quality Control Memory Quick Test Quick Measurement Patient Test Basic Measurement Code Registration Main Menu Caution Product Components Product Overview Power Supply Names of Parts 3 4...

Страница 3: ...correct results may be obtained Product Principle The test contains two monoclonal antibodies specific to analyte The antibodies form a sandwich complex with the analyte in the blood Following removal...

Страница 4: ...oassay is also known as a sandwich assay In this quantitative measurement method by using an antigen a gold nanoparticle labelled detection antibody is combined with an antibody specific to disease ma...

Страница 5: ...components from your nearest dealer For details please contact the store or visit our web site www infopia21 com Optional Components SelexOn test strip SelexOn printer Barcode scanner SelexOn battery...

Страница 6: ...e temperature range 18 32 10 Avoid the electric and magnetic field 11 Use the only SelexOn strip Storage After Use and Maintenance 1 Do not arbitrarily disassemble or alter the device 2 Store the devi...

Страница 7: ...e normal measurement mode SelexOn 11 Power Supply Meter Power Supply 1 Check that the Power Switch is Off 2 Plug the power plug into the wall outlet 3 Connect the External Power Adapter to the Power A...

Страница 8: ...s contact the nearest dealer NOTE Please refer to Troubleshooting if an error message appears CAUTION Ensure that the RFID code is registered when using a new strip box SelexOn 13 Code Registration RF...

Страница 9: ...label blood inlet and measuring window Contaminants may cause an error in the results Step 4 Hold the strip handle and insert the strip into the strip insertion port on the left side of the meter Push...

Страница 10: ...ement time is dis played on the LCD 3 1 2 SelexOn 17 Patient Test Normal Measurement Step 5 The strip marker and code are displayed on the LCD Confirm if the marker and code information on the RFID la...

Страница 11: ...nt Step 9 When the measurement is complete the results appear on the LCD Remove the strip The results are automatically saved Please refer to Device Settings for changing the settings Press Delete to...

Страница 12: ...in Select if the strip reaction time is 20 minutes NOTE Select an appropriate timer after confirming the reaction time on the strip manual If the timer is selected the remaining time is displayed in t...

Страница 13: ...trip The results are automatically saved Please refer to Device Settings for changing the settings Press Delete to delete the results Press Print to print the results on the SelexOn Printer Press Cont...

Страница 14: ...t the top of the list By using the icons the previous or next result can be viewed SelexOn 25 Quick Test Quick Measurement NOTE If the results for CK MB are displayed with an inequality sign such as 4...

Страница 15: ...elected results All Select Selects all results on the current page Displays the test type P Patient Test Q Quick Test Memory Press Memory on the Main Menu and press QC in order to view the QC result l...

Страница 16: ...On Printer Del Deletes the selected results All Select Selects all results on the current page Memory Press Memory on the Main Menu and press QC in order to view the QC result list Press Standard Stri...

Страница 17: ...10 85 humidity Since a standard strip damaged from external shock may affect the results safely store the standard strip CAUTION Do not use an expired standard strip Do not use a standard strip damag...

Страница 18: ...xOn 33 QC Test When to test the standard strip Prior to the sample measurement When the results are deemed to be inaccurate When the QC alarm appears Weekly Monthly NOTE Store the standard strip in it...

Страница 19: ...l representative for technical support Step 6 After completing the test remove the standard strip from the meter Store the standard strip in its storage container while avoiding direct sunlight SelexO...

Страница 20: ...e accuracy of the test strip For details regarding the control solution please refer to the manual provided with the control solution When to run the control solution test Prior to the sample measurem...

Страница 21: ...automatically runs System check Step 6 When the System check is complete the meter automatically runs Heating SelexOn 39 QC Test Step 3 Select a level for the control solution LOW Low concentration H...

Страница 22: ...ay affect the results Please rerun the QC test if FAIL appears If FAIL persists stop using the strip and contact the nearest dealer for service SelexOn 41 QC Test Step 7 When a screen for injecting th...

Страница 23: ...isplay Press or in order to not save the changes Press OK to save the changes CAUTION Please operate the meter after accurately setting the current year month and day If the date is incorrect an error...

Страница 24: ...order varies depending on either the Operator ID or Patient ID on the ID Setup Please refer to ID Input for details BEEP Available BEEP on Touch BEEP LEVEL Set to OFF in order to disable the beep Both...

Страница 25: ...ting for details Device Settings DEVICE Press or in order to not save the changes Press OK to save the changes SelexOn 47 Thermal Printer Auto Print Barcode reader Auto Save Sets the printer Set to OF...

Страница 26: ...ess OK to save the changes ID Input ID Input Settings The SelexOn meter can manage the results by ID through entering either the Operator or Patient ID Press ID Setup in SetUp mode in order to allow e...

Страница 27: ...st by pressing Enter without entering an ID Required must enter the ID An Operator ID input message appears after booting Press ENTER after entering the ID in order to display the Main Menu The OPID i...

Страница 28: ...ON the Thermal Printer 3 After completing the measurement press Print in order to print the results ID Input Patient ID Optional or Required setup Optional performs a test by pressing Enter without e...

Страница 29: ...ode scanner to the Barcode Scanner Port 2 on the rear panel of the meter 2 Turn on the meter Press Device in Setup mode to turn ON the Barcode Reader 3 Press Barcode Reader on the ID Input screen When...

Страница 30: ...may be dam aged Barcode Scanner Optional 4 Scan a barcode with the scanner in Barcode Input Standby mode in order to display the barcode information in the ID input field NOTE Up to 16 characters can...

Страница 31: ...Menu mode CAUTION Do not remove the data cable while running the PC Sync program PC Sync Software PC Sync Software This program transfers the results from the meter to a PC You can download this progr...

Страница 32: ...ntact the nearest dealer Check if the strip code label is con taminated If the label has no prob lem slowly insert the strip again If the problem persists contact the nearest dealer Troubleshooting Er...

Страница 33: ...Level discordance A test strip has used for the stan dard strip test Rerun the test by using the stan dard strip Do not use the standard strip for tests other than the standard strip test Rerun the te...

Страница 34: ...ing and separate the external power adapter and the meter Wipe off dust on the meter with a soft clean cloth Do not use any liquid or gas cleaning agents When cleaning the blood inlet remove the Blood...

Страница 35: ...8kg Storage Temperature 10 35 50 95 Operational Temperature 18 32 64 90 Humidity 10 85 Peripheral Devices Thermal Printer Barcode Scanner Battery 8 4V 4400mAh Data Cable This symbol indicates that the...

Страница 36: ...re Limitation Date of Manufacturer Keep away from sunlight This symbol indicates that the system is type A equipment which provides a particular degree of protection against electric shock Do not reus...