Important Control Solution Test Information
·
Use only the
finetest
TM
Control Solution.
·
Check the expiration date on the control solution vial. DO NOT USE if expired.
·
The
finetest
TM
Normal Control Solution should be stored at room temperature below 86˚F (30˚C).
Do not refrigerate.
·
It is recommended that the meter, control solution, and test strips come to room temperature
(23-25°C) before testing.
·
Shake the control solution, and wipe off the vial tip and reseal tightly after each use.
·
Use only for three months after first opening. Record the discard date on the control solution vial.
Discard after three months.
CAUTION
:
The control solution range printed on the Test Strip vial is for the
finetest
TM
Control Solution only.
It is used to test the meter and test strip performance. It is not a recommended range for your blood glucose level.
[ 9 ]
finetest
TM
INFOPIA
[ 10 ]
How to do the
finetest
TM
Control Solution test
STEP 1: Insert Test Strip
Insert a Test Strip, with the arrow side
“
”
facing up and in the direction of the arrow, insert it into the test port of the Meter firmly. The
meter will automatically turn on and display the code number. Make sure the code number on display matches the code number on the
test strip vial (picture1). If the code number does not match, code the meter correctly. Once completed,
“
Strip-Shape
”
symbol will
appear on the LCD.
STEP 2: Apply Control Solution
Shake the control solution vial well. Remove the cap. Invert the bottle and discard the first drop of the control solution. If large bubbles
are present at the tip of the vial, wipe the bubbles off with a clean tissue before applying the drop of control solution to the test spot. Do
not drop bubbles on the test strip. Apply an entire drop of the control solution onto the test spot at once by gently squeezing the vial to
form a small drop. Take the meter with the test strip inserted and touch the end of the test strip to the hanging drop (picture 2). Be sure
to hold the test strip to the drop until you hear a
“
beep
”
sound.
1
2
3
Checking the System with the Control Solution
Checking the System with the Control Solution