CLOVER A1c Self • 37 •
13. CHECKING THE SYSTEM
3) Gently mix the reagent pack 5~6 times and touch softly the control solution
with the capillary tip, so that the control solution is automatically drawn up
into the sample area. Wait until the sample area is campletely filled with
control solution.
Reagent pack sample area
CAUTION :
Do not mix it too violently. It may make some bubbles.
If you drop the test cartridge and the bubble is arise in the test
cartridge, please wait until the bubble sinks.
ȳ
IMPORTANT :
Once the reagent pack is filled with the control solution,
analysis must begin immediately.
[ adequate ]
[ inadequate ]
CLOVER A1c Self • 36 •
13. CHECKING THE SYSTEM
2) Insert the cartridge when the ‘Cartridge insert’ icon displayed.
HbA1c Control Solution
If external quality control testing is desired, commercial controls from other
vendors may be purchased.
Controls of human whole blood only can be used with the
CLOVER A1c™ Self
.
Please contact your local representative for technical support.
When to run a control test
1) With each new shipment of test cartridges, even if the cartridge are from the same
lot as previously received.
2) With each new lot of test cartridges.
3) If there is concern that test result may be incorrect
WARNING
Use only
CLOVER A1c™ Self
HbA1c Control Solution with the
CLOVER A1c™ Self
system.
1) Open the Cover of
CLOVER A1c™ Self
Analyzer.
How to test with the Control Solution