Mass
Calibration factor SNIF
2
H2
3
M3
4
He
2XL
XL H2
3XL
XL M3
4XL
XL He
LD protocol
Commands 519, 521
ASCII protocol
Command *FACtor:CALSniff or *FACtor:CALSXL for the
current mass
6.3.5 Suppressing gas backgrounds with "ZERO" functions
Unwanted sample gases can be suppressed with ZERO. If ZERO is enabled, the
currently measured leak rate value will be interpreted as carrier gas and subtracted
from all subsequently measured values. The background value suppressed by ZERO
is adjusted automatically if the background changes inside the device. The
background value is automatically adjusted depending on the set ZERO time, except
for filter setting I•CAL, see "Measurement result display with signal filters [
69]".
Activating and
deactivating ZERO
ZERO aktivate/deaktivate
0
On
1
Off
Control unit
Main menu > Function > ZERO > ZERO
LD protocol
Command 6
ASCII protocol
Command ZERO
Setting ZERO mode
Specified the level of the helium background suppressed by ZERO (not with filter
I•CAL).
0
all decades
1
1 – 2 decades
2
2 – 3 decades
3
2 decades
4
3 – 4 decades
5
19/20 of the tracer gas are suppressed
Control unit
Main Menu > Settings > ZERO/filter > ZERO > ZERO mode
LD protocol
Command 410
ASCII protocol
Command *CONFig:DECADEZero
Содержание XL3000flex
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 92: ...9 Decommissioning the measuring instrument INFICON 92 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Страница 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 107: ......
Страница 108: ......