Transpector SPS Operating Manual
4.3.1.4 One Analog Input
One analog input is differential and can handle inputs between 0 to +10 volts and
common mode voltages of 100 volts. See
.
NOTE:
The analog input is supported through FabGuard software.
Figure 4-2 Transpector SPS Aux I/O pinout diagram
4.3.2 Ultra-High Vacuum System
The vacuum system provides low pressure to:
establish a sample flow from the process by pressure difference.
provide low pressure for optimum operation of the sensor.
A dry pumping system comprised of a turbo molecular pump and foreline pump is
used to minimize hydrocarbons in the residual gas background.
The turbo molecular pump provides a high compression ratio between the high
vacuum side and the foreline pump side for all gases, including hydrogen.
Hydrogen is the major residual gas component in ultra-high vacuum. The high
compression ratio for hydrogen is important for producing a low base pressure.
Table 4-3 2 analog inputs
ANALOG INPUT 1
(+) PIN 9
ANALOG INPUT 1
(-)
PIN 10