1 - 18
Transpector MPS Operating Manual
1.17.3 One Analog Input
One analog input is differential and can handle inputs between 0 to +10 volts and
common mode voltages of 100 volts. See
.
NOTE:
The analog input is supported through FabGuard software.
1.18 Installing the Heater Jacket (optional)
The Transpector MPS Heater Jacket Mount Kit (961-433-G1) provided with
Heating Jacket Kit (961-029) is intended to inhibit the movement of the Manifold
Heater. This kit is only for system installations without calibration standards that are
mounted in a vertical orientation. (See
.)
NOTE:
For all other heater installations, discard this kit.
1
Install the Manifold Heater over the tube extension and align the snaps with the
tapped hole in the sensor flange.
2
Align the Internal Heater Bracket so the large hole fits over the snap on the
heating jacket and capture the bracket in the Manifold Heater by securing the
snaps.
3
Using a 7 mm combination wrench, attach the Internal Heater Bracket to the
tapped hole in the sensor flange using the hex head fastener provided in the kit.
4
Attach the manifold heater power connector to the mating connector on the
600-1487 power cable and 068-0433 (120V) or 068-0434 (230V) provided in
the 961-029 heating jacket kit.
Table 1-4 Analog inputs
ANALOG INPUT 1
(+) PIN 9
ANALOG INPUT 1
(-)
PIN 10