![Inficon Sensistor ISH2000 HySpeed Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/inficon/sensistor-ish2000-hyspeed/sensistor-ish2000-hyspeed_operating-manual_2055023040.webp)
Hydrogen leak detector Sensistor ISH2000
-
Operating Manual
40
OM-Sensistor ISH2000-EN-201304
EN
Only pin 2 (Received data), pin 3 (Transmitted data) and pin 5 (Signal ground) are
used. Refer to Figure 8-5 for the wiring diagram.
Fig 8-6. D9 Null modem cable wiring diagram
Selectable printer types
Most PC-printers with serial interface can be connected to the 9-pin printer port.
Parallel (Centronics) interface printer can be used if connected through a serial to
parallel converter.
The port can be set up for the following printer types: PC Printer and Data Dump.
No printer
Printer output disabled. Incoming communication is enabled. Sensistor ISH2000
listens for incoming data but will not print/send test results.
PC printer (with serial interface)
The
PC Printer
option can be used to print data on most standard PC printer with
serial interface. Parallel interface printers can be used if connected through a serial
to parallel converter (see below).
Note:
The output format has been chosen to be as simple as possible to ensure that
most printers will accept it. Therefore, the printer output does not use any
flow control. This means that some printers may delay printing until the input
buffer is full or a pre-defined timeout has elapsed.
Table 8-5. Communication specifications.
5
SG
Signal ground
6
(DSR)
Not used
7
(RTS)
Not used
8
(CTS)
Not used
9
+7 VDC
Power feed USB/RS232 converter
Pin
Signal
Comments
1
2
3
4
5
6
7
8
9
9
8
7
6
5
4
3
2
1
CD
CD
RxD
RxD
TxD
TxD
DTR
DTR
SG
SG
DSR
DSR
RTS
RTS
CTS
CTS
R1
R1
D9 Female
D9 Female
Setting
Value
Data rate
1200 baud
Data bits
8
Stop bits
1
Parity
None
Flow control
None