10 - 49
IP
N 07
4-
50
5-
P1
E
IC6 Operating Manual
10.4.33 Remote General Action
Command
= RG <Command ID> (<Action Value>)
Command ID
.
Action Value
= <Byte>… <Byte> | <Integer>…<Integer>|<Float>…<Float>
See
for commands that require an action value.
Response
= None (Just header and trailer)
Table 10-15 Remote general
RG
Command
ID
Meaning
Action Value
Notes
0 (0x00)
Start
Duplicates the equivalent front panel action
1 (0x01)
Stop
Duplicates the equivalent front panel action
2 (0x02)
Reset
Duplicates the equivalent front panel action
3 (0x03)
Remote Lock On
No parameters can be updated from the front panel
"R Lock" displayed
4 (0x04)
Remote Lock Off
Cancel RG 3 command
5
(0x05)
Set Logic
Statement
<Statement
Number> to True
<Byte>
Range is 1-100.
Only does something if logic statement event "Computer
Control" set. Otherwise, ignored.
6
(0x06)
Clear Logic
Statement
<Statement
Number> to False.
<Byte>
Range is 1-100.
Only does something if logic statement event "Computer
Control" set. Otherwise, ignored.
7
(0x07)
Crystal Fail Inhibit
On
<Byte>
Range is 0-8 (0 = All)
"Crystal Fail" output always set to FALSE, even during
crystal fail
8
(0x08)
Crystal Fail Inhibit
Off
<Byte>
Range is 0-8 (0 = All)
"Crystal Fail" output has normal meaning
9
(0x09)
Set date (Month,
Day, Year) or (Day,
Month, Year)
Depending on
General parameter
17 (Date Format).
<Integer>
<Integer>
<Integer>
Range as follows:Month = 1 to 12
Day = 1 to 31
Year = 2000 to 2099
10
(0x0a)
Set time (Hour,
Minute, Second)
<Integer>
<Integer>
<Integer>
Range as follows Hour = 0 to 23
Minute = 0 to 59
Second = 0 to 59
Содержание IC6
Страница 1: ...O P E R A T I N G M A N U A L IC6 Thin Film Deposition Controller IPN 074 505 P1E Cover Page ...
Страница 2: ......
Страница 8: ......
Страница 20: ...TOC 12 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 42: ...1 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 134: ...6 6 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 148: ...8 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 170: ...9 22 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 236: ...11 2 IPN 074 505 P1E IC6 Operating Manual Figure 11 1 The essential advantage of multi point deposition sensing ...
Страница 268: ...13 2 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 312: ...15 40 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 338: ...A 10 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...
Страница 342: ...1 4 IPN 074 505 P1E IC6 Operating Manual This page is intentionally blank ...