HAPSITE Smart Operating Manual
11.3.2 SituProbe
To create a SituProbe method, choose the
SituProbe Sample Input Device
. This
will bring up another temperature setting window for setting the SituProbe oven and
transfer line temperatures. See
.
Figure 11-9 SituProbe Parameters
SituProbe Temperatures (C)
Oven
. . . . . . . . . . . . . . . . . . . . . . . . The target temperature the system will
achieve in the Oven before starting data
acquisition. This setting is not available when
the SituProbe or Headspace is disabled.
Transfer Line
. . . . . . . . . . . . . . . . . . The target temperature the system will
achieve on the Transfer Line before starting
data acquisition. This setting is not available
when the SItuProbe or Headspace is
disabled.
Probe
. . . . . . . . . . . . . . . . . . . . . . . . The target temperature the system will
achieve in the SituProbe before starting data
acquisition.
11.4 Inlet Page
NOTE:
The Inlet page is not used (or displayed) when creating survey methods.
shows the default settings for the Inlet page for a GC/MS Full Scan
Method. From this page, the Inlet States, GC Temperature Profiles and Run Time
are integrated and synchronized to form the Inlet Component of the Method.
Changing one item on this page often affects others and that effect is shown in the
profile, at the bottom of the page, for clarity. There is a direct connection on this
page between the start time of the Standby Inlet Event, the total time of the
Temperature Profile, and the Run Time. Changing a time related to one of these
items will automatically change a related time in the other items, in most cases.
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......