INFICON
Operation CU1000 | 6
CU1000-operating-instructions-jina54en1-06-(1901)
19 / 36
Pinpoint: The sound of the acoustic signal changes its frequency within a specific
range of leak rates. Range: A decade below the selected trigger threshold up to one
decade above. The sound keeps at a constant low and a constant high frequency
below and above this range, respectively.
Trigger: If the selected trigger threshold is exceeded, a two-pitch signal sounds.
Control unit
Settings > Set up > Control unit > Audio > Audio alarm mode
Behavior with warnings or error messages: If the touch screen shows a warning or an
error, then a two-pitch signal sounds simultaneously.
Automatic switch off of
the touch screen
The touch screen can be switched off automatically after a specific time without any
operation to save energy.
30 s
1 min
2 min
5 min
10 min
30 min
1 h ∞ (=never)
Control unit
Settings > Set up > Control unit > Energy > Display off after
6.3.2 Operator types and authorizations
There are four different operator types that are distinguished by different
authorizations. The integrator is registered ex works.
Additional operators can be registered. The following table shows options for
individual operator types to register new operator types.
Operator registration
Viewer
Operator
Supervisor
Integrator
-
Operator
Viewer
Supervisor
Operator
Viewer
Integrator
Supervisor
Operator
Viewer
For the types "Integrator", "Supervisor" and "Operator", a four-digit PIN must be
assigned during registration (0000 ... 9999). "0000" is assigned to all operators ex
works.
If an operator keeps the pin "0000", this operator will always be registered is during
the start up of the system (without PIN query).
A key-operated switch can be used in addition to a PIN if an I/O module is connected.
The key-operated switch is connected to the I/O module via three digital inputs (see
operating instructions of the LDS3000).
The following table shows the authorizations of individual operator types.
Содержание CU1000
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 30: ...7 Decommissioning the measuring instrument INFICON 30 36 CU1000 operating instructions jina54en1 06 1901 ...
Страница 32: ...8 Appendix INFICON 32 36 CU1000 operating instructions jina54en1 06 1901 8 2 RoHS ...
Страница 33: ...INFICON Appendix 8 CU1000 operating instructions jina54en1 06 1901 33 36 ...
Страница 34: ...8 Appendix INFICON 34 36 CU1000 operating instructions jina54en1 06 1901 ...
Страница 35: ......
Страница 36: ......