4 Description
4.1 Function
The IRwin is a portable methane leak detector for gas pipeline survey. It measures
following gases depending on the leak detector version:
• Methane
• Carbon dioxide
• Ethane
• Propane
• Butane
• Oxygen
• Hydrogen sulfide
• Carbon monoxide
There are several probes available for the leak detector:
• The carpet probe is used for surveying paved or smooth surfaces. Pull or push the
carpet along the surface.
• Use the bell probe for measuring gas at specific points. Press the bell firmly down
to pull gas through the surface.
• The hand probe is used for checking features on or in buildings. The hand probe
also constitutes the handle for the carpet probe and bell probe.
• The Bar Hole probe is used to sample air from drilled holes. Push into hole and
tighten screw to expand rubber before connecting sample line.
• The optional Hand Probe Flexible Extension can be connected to the hand probe
or rod for easier access.
• The Swan Neck Probe is used on uneven surfaces and in bushes etc.
• The Flexi Bell is used for simple survey including meters and piping. The bell
adapts to pipe geometry.
INFICON single-cell
wide range technology
(patent pending)
IRwin methane leak detector uses the unique INFICON single-cell wide range
technology for detecting all concentrations from 1 ppm to 100 Vol% methane. This
technology is very sensitive to the ethane content of natural gas. Therefore, large
deviations can occur at higher methane concentrations if the ethane concentration is
not compensated for. The instrument has built-in routines for ethane compensation.
The actual ethane concentration can be determined using the IR Ethane Analysis
function. This can be used to automatically set the compensation concentration. If
preferred you can instead set the ethane content manually. The safety certified modes
Ex and Ex Tox are not affected by this phenomenon. See "IR Ethane Analysis and
Содержание 580-000
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 96: ...8 Decommissioning the leak detector INFICON 96 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 99: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 99 128 ...
Страница 101: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 101 128 ...
Страница 102: ...9 Certificates INFICON 102 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 103: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 103 128 ...
Страница 105: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 105 128 ...
Страница 106: ...9 Certificates INFICON 106 128 IRwin Operating instructions mina66en1 15 1904 9 3 Certificate for software for Ex ExTox ...
Страница 110: ...9 Certificates INFICON 110 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 111: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 111 128 ...
Страница 113: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 113 128 ...
Страница 114: ...9 Certificates INFICON 114 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 115: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 115 128 ...
Страница 116: ...9 Certificates INFICON 116 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 117: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 117 128 ...
Страница 118: ...9 Certificates INFICON 118 128 IRwin Operating instructions mina66en1 15 1904 9 8 CE Declaration of Conformity ...
Страница 119: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 119 128 ...
Страница 120: ...9 Certificates INFICON 120 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 121: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 121 128 ...
Страница 122: ...9 Certificates INFICON 122 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 123: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 123 128 ...
Страница 126: ...Index INFICON 126 128 IRwin Operating instructions mina66en1 15 1904 ...
Страница 127: ......
Страница 128: ......