Environment Requirements

Room Temperature:

15 - 30°C (59 - 86°F)

Relative Humidity:

40 - 76% RH (noncondensing over operating temperature range)

Altitude:

Maximum altitude 2424 m (8000 ft) above sea level.

Ambient Light:

Room lighting should not exceed 450 lux (150 ft-candles) at the
PROCESSOR. The room must be capable of going completely dark
when loading film into the PROCESSOR.

Heat Output:

2200 Btu/hour

Ventilation:

• Volume - full load, 2,100 L/min (75 ft

3

/min), maximum, 24 hours per day

• 66°C (150°F) maximum
• EXHAUST DUCT from the building with an Adjustable Air Gap - 0.76
- 2.54 mm (0.03- 0.10 in.) of water
• 10.2 cm (4 in.) DUCT 0.25 - 1.02 mm

Check local codes for venting requirements.

• If venting is not correct, fumes will corrode the equipment. Do not install the PROCESSOR or
ACCESSORIES if the venting is not correct.
• If the PROCESSOR is installed with an AUTOFEEDER or FEED TABLE is located in the darkroom and the
PROCESSOR is installed in a daylight location, the darkroom must have adequate pressure to ensure a
positive airflow from the FEED to the DRYER. This should avoid condensation problems.
• The airflow is correct when the fumes are flowing out the PROCESSOR through the EXHAUST HOSE.
• If the ventilation is to be connected to the PROCESSOR, measure negative static pressure in the
EXHAUST DUCT. See on the page x.

Plumbing Requirements

• All plumbing requirements must comply with local and national codes. Do not use IRON PIPES.
• All DRAIN material must be made of chemically resistant, non-corrosive material. Use PVC or the
equivalent.

Water Supply 

• Temperature

– 10° - 60°C (64° - 140°F)
– If the temperature of the water supply is higher than 60°C (140°F), install a
WATER CHILLER.
– A tempered water supply is recommended for cleaning the PROCESSOR and for
mixing chemicals manually.
• Filtered
– 50-micron WALTER FILTERED is recommended in the input water line
• Flow Volume
– 2.5 L/min
• Pressure
– 3 - 10 bars
• Location:
– Accessible to both the PROCESSOR and the REPLENISHMENT TANKS

HOSES 

2-way magnetic VALVE with 3/4” HOSE connection using a DVGW system device or

PIPE device

DRAIN 

• Size: 32 mm (5.4 in.) HOSE connection

• Minimum Diameter - 7.6 cm (3 in.) with no obstructions
• Distance from PROCESSOR - 1.5 m (60 in.)
• Height from floor - Top of the DRAIN or DRAIN CONTAINERS must be lower than
the bottom of the PROCESSOR.
• Capacity
– 1 L/min (1/4 gal/min) during normal operation
– 2.85 L/min 3/4 gal/min) for draining all three solutions together
– 0.95 L/min (1/4 gal/min) if each solution is drained separately

Содержание M43ic

Страница 1: ...page 1 9 document no X1 25_4 pdf NDT Film Processors type produced by Colenta To be used to process all standard NDT Films PRE INSTALLATIONMANUAL NDT FILM PROCESSOR 01 2013 TK Industrex M43ic ...

Страница 2: ...loor or wall drain may be used which should include an anti syphon system Note Do not use brass or copper in the drain lines The minimum diameter of the drain lines should be 40 mm Electrical For operation a seperate 16 Amp fused socket with earth leakage protection is required for further information consult the processor datasheet Replenishment Each replenisher storage tank has a 30 litre capaci...

Страница 3: ...n the fumes are flowing out the PROCESSOR through the EXHAUST HOSE If the ventilation is to be connected to the PROCESSOR measure negative static pressure in the EXHAUST DUCT See on the page x Plumbing Requirements All plumbing requirements must comply with local and national codes Do not use IRON PIPES All DRAIN material must be made of chemically resistant non corrosive material Use PVC or the e...

Страница 4: ...ocument no X1 25_4 pdf DEV FIX 3 4 Watersupply Drain DEV FIX 2x30L Replenishment containers are supplied with the processor Water Supply Drain connection and Replenishment Exhaust port see more on page 6 OPTIONAL ...

Страница 5: ...3 2 90 45 001 Kugelhahn 3 4 In Line Tap 3 3 90 45 002 Doppelnippel 3 4 3 4in straight fitting 8 4 90 45 003 T Stück 3 4 T Fitting 3 4in 1 5 90 45 004 Bogen 3 4 90 3 4in Elbow 90 2 6 90 45 005 Reduzierung Male Female fitting 2 7 90 45 007 Holländer Filter inlet outlet fitting 2 8 90 45 009 Dichtung Flat washer 2 9 90 45 027 Filtergehäuse Fiter Housing 1 10 90 45 035 Filterpatrone Filter Cartridge 1...

Страница 6: ...r Air Meter can then be inserted through the opening When measuring negative air the tube tip opening should be pointed in the direction of airflow away from the processor The processor must be de energized when making air measurements The air meter should be held in the vertical position to assure the greatest accuracy The meter tubing must not be kinked It is most important that negative airflow...

Страница 7: ...page 7 9 document no X1 25_4 pdf recommended Installation Free standing Dark Stop cock 50cm Positive Air Pressure ...

Страница 8: ...page 8 9 document no X1 25_4 pdf recommended Installation Processor Darkroom Exit to light Dark Daylight Stop cock 50cm Positive Air Pressure ...

Страница 9: ...page 9 9 document no X1 25_4 pdf recommendedInstallation Infeed in Darkroom processor in daylight Dark Daylight Stop cock 50cm Positive Air Pressure ...

Отзывы: