NDT - Filmprocessor

The processor is designed in such a way that only a cold water supply, drain and power connection are
necessary. The plumbing is carried out according to DIN 1986/1988 and must comply with local plumbing
codes.The cold water supply must have a shut-off valve / tap system connected with a ¾“ outlet (washing
machine connection) to the processor.

Make sure that the water supply tap is always accessable as it must be opened before work and closed
after work. By means of the pressure/flow regulator (integrated into the water solenoid valve of the
processor) the flow rate is controlled to 2.5Lt/min assuming a water supply pressure of 0.3-10 bar.
It is recommended to install a second water supply outlet (with shut-off tap) and with approx. 2.5 mtr/8ft of
hose for rack cleaning purposes.

The drain tubes of the processor can be drained separately or together, according to local regulations
and health/safety requirements. The hose connections from the processor to the outlet (drain) are
enclosed. The 

fixer

 can be collected separately in a plastic container (storage tank) or directly connected

to a silverrecovery unit. The 

developer

 overflow can be collected in a plastic container. In order to avoid

a backwash of the drained, used chemicals, the drain hoses should be free of bends and with a constant
fall. The drain must be correctly ventilated!

A floor or wall drain may be used which should include an anti syphon system.

Note: Do not use brass or copper in the drain lines.

The minimum diameter of the drain lines should be 40 mm.

Electrical:

For operation, a seperate 16 Amp fused socket with earth leakage protection is required.
(for further information, consult the processor datasheet)

Replenishment:

Each replenisher storage tank has a 30 litre capacity and should be located on ground level with easy
access - please take notice of the required minimum wall clearence for servicing and cleaning

Automatic Cooling:

 The processor electronics will automatically detect over temperature developer

conditions and then activate a cold water cooling system (Chiller). Where a Chiller is not used, the
temperature of the incoming „mains“ cold water supply should be between 7 - 15 C in order for the
cooling system to operate efficiently.

Processor ventilation:

 The processor is  supplied with an exhaust port located at the feed end of the

processor. During installation, the safety cover of the port should be removed by a qualified person and
the open port must then connected to the main exhaust facility at the installation site (with sufficient power
to ventilate the warm exhaust air away.

Where the processor is installed in a „through wall“ location whereby an auto feeder or feed table is
positioned in a darkroom and the main body of the processor is in daylight, it is important that the
darkroom is pressurised to ensure a positive airflow from feed to dryer thus avoiding condensation related
problems.
...........................see more about  this on page 6

Industrex M43ic

Содержание M43ic

Страница 1: ...page 1 9 document no X1 25_4 pdf NDT Film Processors type produced by Colenta To be used to process all standard NDT Films PRE INSTALLATIONMANUAL NDT FILM PROCESSOR 01 2013 TK Industrex M43ic ...

Страница 2: ...loor or wall drain may be used which should include an anti syphon system Note Do not use brass or copper in the drain lines The minimum diameter of the drain lines should be 40 mm Electrical For operation a seperate 16 Amp fused socket with earth leakage protection is required for further information consult the processor datasheet Replenishment Each replenisher storage tank has a 30 litre capaci...

Страница 3: ...n the fumes are flowing out the PROCESSOR through the EXHAUST HOSE If the ventilation is to be connected to the PROCESSOR measure negative static pressure in the EXHAUST DUCT See on the page x Plumbing Requirements All plumbing requirements must comply with local and national codes Do not use IRON PIPES All DRAIN material must be made of chemically resistant non corrosive material Use PVC or the e...

Страница 4: ...ocument no X1 25_4 pdf DEV FIX 3 4 Watersupply Drain DEV FIX 2x30L Replenishment containers are supplied with the processor Water Supply Drain connection and Replenishment Exhaust port see more on page 6 OPTIONAL ...

Страница 5: ...3 2 90 45 001 Kugelhahn 3 4 In Line Tap 3 3 90 45 002 Doppelnippel 3 4 3 4in straight fitting 8 4 90 45 003 T Stück 3 4 T Fitting 3 4in 1 5 90 45 004 Bogen 3 4 90 3 4in Elbow 90 2 6 90 45 005 Reduzierung Male Female fitting 2 7 90 45 007 Holländer Filter inlet outlet fitting 2 8 90 45 009 Dichtung Flat washer 2 9 90 45 027 Filtergehäuse Fiter Housing 1 10 90 45 035 Filterpatrone Filter Cartridge 1...

Страница 6: ...r Air Meter can then be inserted through the opening When measuring negative air the tube tip opening should be pointed in the direction of airflow away from the processor The processor must be de energized when making air measurements The air meter should be held in the vertical position to assure the greatest accuracy The meter tubing must not be kinked It is most important that negative airflow...

Страница 7: ...page 7 9 document no X1 25_4 pdf recommended Installation Free standing Dark Stop cock 50cm Positive Air Pressure ...

Страница 8: ...page 8 9 document no X1 25_4 pdf recommended Installation Processor Darkroom Exit to light Dark Daylight Stop cock 50cm Positive Air Pressure ...

Страница 9: ...page 9 9 document no X1 25_4 pdf recommendedInstallation Infeed in Darkroom processor in daylight Dark Daylight Stop cock 50cm Positive Air Pressure ...

Отзывы: