Workplace requirements
•
Location: Indoor use only. Any outdoor area where the equipment is to be
located should meet all the conditions below.
•
Operation environment: 10~40
℃
(50~104
℉
), 30~75%RH, 70~106kPa
•
Adapter: Power Input 100~240V, 50/60Hz, 1.2A
Power Output DC 12V, 3.33A
Note on Unpacking & Assembling
(1) Place the BSM370 on level ground.
(2) Do not move by holding the head bar or operation part with LCD.
2. Raise the stand of the BSM370
❶
Raise the stand of the BSM370 carefully until it stands upright.
Tighten the joint screw connecting stand body and foot plate.
4. Balance the BSM370
Use the bubble level indicator on the BSM370 to check it is level. Use the
leveling screw underneath the foot plate of the BSM370 if it is not level.
❷
Turn on the power switch. The head bar will move up automatically.
Head bar of BSM370 will be automatically went up after turning the
device on.
3. Fixed the head bar
Unscrew the fixed screw at the head bar
part and uplift it by 90 degree.
Screw tightly.
1. Open the box
Open the packing box and remove the wrapping.
Take out all contents.
5. Power Connection
❶
Connect the power adapter for the BSM370 which is placed on the
back of the foot plate.
❷
After the upper part of the BSM370 is fixed,
close the joint cover.
Bubble level indicator
Leveling screw
<Lowering>
<Raising>
<Un-leveled state>
<Leveled state>
screw
Remote control device is provided with thermal printer which is attached on.
NOTE
Save the wrapping material after unpacking in the event of relocation.
NOTE
Do not lift the head bar to prevent any malfunctioning.
WARNING
Make sure you use the power adapter and main cable provided by
InBody.
WARNING
If the equipment is not horizontally level, adjust the leveling screw on the
opposite side of the bubble to place it at the center of the level indicator.
NOTE
In case of a loose connection between the stand body and the foot plate,
measurement process cannot be performed properly.
Fix the stand perpendicular on the foot plate.
CAUTION
Please check the height from the ground to the ceiling before installing
BSM370. It should be at least as tall as 2156mm ( 84.9in.).
BSM370’s height is 2126mm (83.7in.).
CAUTION
When pressure or weight is applied to the foot plate during self-calibration,
message appears. Turn the BSM370 off and on again after removing
the material from the foot plate. Please do not apply pressure or weight to the
foot plate during self-calibration.
Be careful not to put any weight on the foot plate. If you step on the foot plate
or put a heavy object on the foot plate, it will affect the calibration and
measurements will not be accurate.
CAUTION
NOTE
Quick Reference
❷
❸
❹
❺
❼
❻
❶
❾
❿
❽
Included Items
❶
BSM370
❷
Quick Reference
❸
User’s manual CD
❹
Adapter (12V, 3.33A) 1EA
❺
Power cord
(AC 250V, 10A, 1.8m( 5ft. 10in.)) 1EA
❻
Screw
❼
Joint cover
Optional
❽
Thermal printer 1EA , Thermal printer base
Screw (M4
×
8) 6EA, Cable (95cm (3ft. 1in.))
❾
Remote control device, Cable (2m (6ft. 7in.),10cm (4in.))
❿
Cable - Y type