28
(8) Completing the Measurement
❶
If the pulse is not detected or unstable, open the exhaust valve and exhaust the air compressed
in the Cuff completely.
❷
When the pressure drops below 20mmHg, the Heart Rate is displayed as a number in the
Pressure / Pulse Rate Display on the screen.
The heart rate is measured when the pressure is reduced at a rate of 2 to 4 mmHg/sec.
If the pulse is not detected or unstable, ‘- - -’ is displayed on the screen.
* Please review ‘B. Measurement related - Pulse rate has not been measured’ on page 56.
(9) Turning the Power Off
❶
press the Button to switch off the power.
MARK Button
Clockwise Rotation: Increase Exhaust Speed
Counter-clockwise Rotation: Decrease Exhaust Speed
Pressing: Enable Mark Function
Exhaust Vale Closed
Wheel Switch
Exhaust Valve Open
PUSH
(7) Depressurization
❶
Rotate the Exhaust Valve counterclockwise to open. Or, rotate the wheel switch to adjust
the exhaust speed. Proceed with blood pressure measurement at a proper exhaust rate.
❷
If the button on the Main Body is pressed at a certain pressure value, measured
blood pressure value will be displayed accordingly.
* The Mark Function can show up to 5 results. If there is more than 5 measurements, the previous value will be
erased automatically.