SETTING
3. When finished viewing memory, you
may use the device as explained in #3
of Quick Start (pg.5).
Deleting Memories
Data stored in the memory will disappear
after replacing the batteries.
USE
Measurement Consideration
CAUTION
1. Forehead site must be dry and clean,
free of sweat and cosmetics prior to
taking measurement.
2. Hair should be held back from forehead
site before taking measurement.
Aim the probe center at the temple (around
2 cm above eyebrow tail) or object surface.
Keep the probe away from skin or object
surface around 5cm.
USE
Taking a Measurement
1. Press and release the POWER Button
to turn on the device.
2. All symbols should
appear on the display.
3. Aim the Probe toward the forehead or
object, approximately 5cm from surface,
then press and release the POWER
Button to record the measurement.
4. The measurement is complete when
you hear “Di Di”-“Di Di” sounds.
5. You can take next measurement
instantly.
6. The temperature and back light color
shows on the display.
7. The color of the back light is determined
by measurement reading. Please refer
to Table 2.
Table 2
Temperature Reading
Back Light
Color
°C
°F
T
≦
37.1
T
≦
98.9
Green
37.2
≦
T
≦
37.4 99
≦
T
≦
99.4
Yellow
T
≧
37.5
T
≧
99.5
Red
8. In Forehead Mode Only: The device
will continuously beep a fever alarm if
the temperature reading is over 37.5°C
(99.5°F).
USE
9. The device will automatically power off
after 60 seconds of no use.
CLEANING
1. Use a cloth dampened with water or
neutral detergent to clean and dry the
device and Lens.
2. Do not use thinner, benzene, and other
harsh cleaners for cleaning the device.
3. Do not immerse the device in water
or liquid of any kind. The device is
not water proof.
CARE AND STORAGE
1. DO NOT drop or crush the device. It is
not shock proof.
2. DO NOT dismantle or attempt to modify
the device.
3. DO NOT store the device in an extreme
environment with direct sunlight, very
high or low temperatures or in a place
with high moisture/humidity or dust.
4. Please contact your local agent or
distributor if the device does not
function as described in this manual.
PRODUCT SPECIFICATION
Measurement Sites
Forehead and Object
Measurement
Distance
3 ~ 8cm
Measurement Time
1 second
Display Resolution
0.1°C (0.1°F)
Measurement Scale
°C / °F
Measurement Range
Forehead mode :
32.2 ~ 43.3°C (90 ~ 110°F)
Object mode :
0 ~ 100°C (32 ~ 212°F)
Accuracy
Forehead mode :
32.2 ~ 43.3°C (90 ~ 110°F)
±0.2°C(0.4°F)
Object mode :
0 ~ 100°C (32 ~ 212°F) ± 1°C (2°F)
Memory
10 sets
Auto Shut Off
After 60 seconds
Operation
Environment
10 ~ 40°C (50 ~ 104°F)
15 ~ 95%RH (non condensing)
Storage Environment -25 ~ +55°C (-13 ~ +131°F)
15 ~ 95%RH (non-condensing)
Power Source
2 AAA 1.5Volt batteries
Battery Life
Approx. 1000 measurement cycles
Dimension
138 x 90 x 45 mm
( 5.43 x 3.54 x 1.77 in )
Weight
132g (4.66oz) including batteries
Warranty
1 year
Classification
Internally powered equipment
Type BF Applied Part
This device complies with all requirements
of ASTM E1965-98 Standard Specification
for Infrared Thermometers for Intermittent
Determination of Patient Temperature, EN
12470-5 Clinical thermometers - Part 5 :
Performance of Infra-red ear thermometers
(with maximum device)