SETTING

3. When  finished  viewing  memory,  you 

may use the device as explained in #3 

of Quick Start (pg.5).

Deleting Memories

Data stored in the memory will disappear 

after replacing the batteries.

USE

Measurement Consideration

 

CAUTION

1. Forehead site must be dry and clean, 

free of sweat and cosmetics prior to 

taking measurement.

2. Hair should be held back from forehead 

site before taking measurement. 

Aim the probe center at the temple (around 

2 cm above eyebrow tail) or object surface. 

Keep the probe away from skin or object 

surface around 5cm.

USE

Taking a Measurement

1. Press and release the POWER Button 

to turn on the device. 

2. All  symbols  should 

appear on the display.

3. Aim the Probe toward the forehead or 

object, approximately 5cm from surface, 

then  press  and  release  the  POWER 

Button to record the measurement.

4. The measurement is complete when 

you hear “Di Di”-“Di Di” sounds.

5. You  can  take  next  measurement 

instantly.

6. The temperature and back light color 

shows on the display.

7. The color of the back light is determined 

by measurement reading. Please refer 

to Table 2.

 Table 2

Temperature Reading

Back Light 

Color

°C

°F

T

37.1

T

98.9

Green

37.2

T

37.4 99

T

99.4

Yellow

T

37.5

T

99.5

Red

8. In Forehead Mode Only: The device 

will continuously beep a fever alarm if 

the temperature reading is over 37.5°C 

(99.5°F).

USE

9. The device will automatically power off 

after 60 seconds of no use.

CLEANING

1. Use a cloth dampened with water or 

neutral detergent to clean and dry the 

device and Lens.

2. Do not use thinner, benzene, and other 

harsh cleaners for cleaning the device.

3. Do not immerse the device in water 

or liquid of any kind. The device is 

not water proof.

CARE AND STORAGE

1. DO NOT drop or crush the device. It is 

not shock proof.

2. DO NOT dismantle or attempt to modify 

the device.

3. DO NOT store the device in an extreme 

environment with direct sunlight, very 

high or low temperatures or in a place 

with high moisture/humidity or dust.

4. Please  contact  your  local  agent  or 

distributor  if  the  device  does  not 

function as described in this manual.

PRODUCT SPECIFICATION

Measurement Sites

Forehead and Object

Measurement 

Distance

3 ~ 8cm

Measurement Time

1 second

Display Resolution

0.1°C (0.1°F)

Measurement Scale

°C / °F 

Measurement Range

Forehead mode : 

32.2 ~ 43.3°C (90 ~ 110°F)

Object mode : 

0 ~ 100°C (32 ~ 212°F)

Accuracy

Forehead mode : 

32.2 ~ 43.3°C (90 ~ 110°F) 

±0.2°C(0.4°F)

Object mode : 

0 ~ 100°C (32 ~ 212°F) ± 1°C (2°F)

Memory

10 sets

Auto Shut Off

After 60 seconds

Operation 

Environment

10 ~ 40°C (50 ~ 104°F)

15 ~ 95%RH (non condensing)

Storage Environment  -25 ~ +55°C (-13 ~ +131°F)

15 ~ 95%RH (non-condensing)

Power Source

2 AAA 1.5Volt batteries

Battery Life

Approx. 1000 measurement cycles

Dimension

138 x 90 x 45 mm 

( 5.43 x 3.54 x 1.77 in )

Weight

132g (4.66oz) including batteries

Warranty

1 year

Classification

Internally powered equipment

Type BF Applied Part

This device complies with all requirements 

of ASTM E1965-98 Standard Specification 

for Infrared Thermometers for Intermittent 

Determination of Patient Temperature, EN 

12470-5 Clinical thermometers - Part 5 : 

Performance of Infra-red ear thermometers 
(with maximum device)

Отзывы: