SETTING
Changing Modes/Settings
The device has four modes/settings: 1.
Forehead mode 2. Object temperature
mode 3. Beeper On/Off setting 4. Offset
setting
The default mode is Forehead mode.
1. Press and release the POWER Button
to turn on device.
2. Press and hold the Memory/Mode
Button firmly for about 3 seconds.
3. Continue to press and release the
Memory/Mode Button to switch between
the four modes/settings.
Forehead mode
This mode is intended to measure the
temperature of forehead.
1. When switching between the four
modes/settings, simply stop pressing
the button when the screen show
.
2. The device is now ready to measure
forehead temperature.
Refer to USE section on page 12 under
Measurement Consideration
. Carefully
read and follow these instructions.
SETTING
Object temperature mode
This mode is intended to measure the
temperature of object surface.
This mode
is not intended for determining human
body temperature.
1. When switching between four modes/
settings, simply stop pressing the button
when the screen shows
.
2. The device is now ready to measure the
surface of object temperature.
Beeper On/Off
The default for beeper is ON. You may
switch it off to avoid disturbing the patient
or others.
1. When switching between the four
modes/settings, simply stop pressing
the button when the screen shows
.
2. After the
disappears, the device
will show
on
. Press and release the
Memory/Mode Button to show
OFF
.
Press the POWER Button to confirm
sound is off.
3. To turn the sound back on, repeat step
1 above.
4. After the
disappears, the device
will show
OFF
. Press and release the
SETTING
Memory/Mode Button to show
on
.
Press the POWER Button to confirm
sound is on.
Offset Setting
The device is capable to convert
measured forehead temperature reading
to oral temperature automatically. You
can apply this mode as a reference
temperature to other body sites such as
rectal, axillary, and auricular. Please use
other thermometer to measure your rectal,
axillary or auricular temperature to confirm
the temperature difference. Offset default
is zero (off).
Table 1
Body Sites Normal Temperatures
Oral
35.5°C ~37.4°C (95.9°F ~99.3°F)
Rectal
36.2°C ~37.9°C (97.1°F ~100.2°F)
Axillary
35.0°C ~37.0°C (95°F ~98.6°F)
Auricular
35.7°C ~37.9°C (96.2°F ~100.2°F)
1. When switching between the four
modes/settings, simply stop pressing
the button when the screen shows
.
2. Press the Memory/Mode Button to
start offset alignment when symbol
disappears.
3. Press the POWER Button to increase
offset within range from +1°C to -1°C
(+2°F to -2°F).
SETTING
4. It increases 0.1°C (0.2°F) by pressing
the offset manually.
5. Press and release the Memory/Mode
Button to turn the offset off.
6. The device will beep twice indication it
is ready to begin a new measurement.
Setting °C or °F Scale
1. You may only set/reset the °C or °F
scale when the device in Forehead
mode or Object mode.
2. After device is powered on, press and
release the Memory/Mode button, then
immediately press and release the
Power Button to switch the temperature
scale.
3. Device is now ready to measure a
temperature.
4. Repeat step 2 above to switch back and
forth.
Memory Functions
Recalling Memories
1. Memory recall can only be used when
the device is in Forehead mode or
Object mode.
2. After device has been powered on,
simply press and release the Memory/
Mode Button to show readings stored in
memory.