1-800-446-7488
33
www.in-situ.com
If you are using a custom standard, the app will not automatically detect it. Instead, a field will
appear in which you can enter the appropriate value. Select
Set User Defined
to begin
calibrating with the new value. If you are not using a custom standard and the app does not
automatically detect the standard, perform the sensor cleaning and maintenance procedure,
then select
Retry Auto Detect
.
13. After the calibration is stable, select
Accept
.
14. The calibration values are applied to the sensor and appear on screen. You can view a full calibration report for all
sensors, or select
Done
to return to the Calibration Menu.
15. Rinse the sensors and restrictor with DI water.
Calibrate the Level Sensor
The factory calibration of the level sensor is very accurate. In-Situ does not recommend calibrating the Level sensor unless
your SOP specifically requires you to do so.
1. From the main menu, select
Connected Instrument
.
2. Select
Calibrations
.
3. From the Calibrations menu select
Level
.
4. Make sure that the pressure sensor is open to air and not submersed in water.
5. Select
Next
.
6. After the calibration is stable, select
Accept
.
Calibrate the pH Sensor
You can calibrate the pH sensor with either a 1-point, 2-point, or 3-point process.
1. From the main menu, select
Connected Instrument
.
2. Select
Calibrations
.
3. From the Calibrations menu select
pH
.