Calibrate the Conductivity Sensor
1. From the main menu, select
Connected Instrument
.
2. Select
Calibrations
.
3. From the Calibration Menu, select
Conductivity
.
4. Make sure the restrictor is in calibration mode.
5. Take the calibration solution and pour 10-20 mL (1/2” above the sensor face) on top of the sensors to perform the
first rinse.