50
TROLL
9500
Operator’s Manual
0095110 rev. 007 01/09
For
DEPTH:
1 Select Pressure in Parameters list, select Edit
2
Display Mode:
Select Depth, click Next.
3 Select Depth/Level Conversion: Specific Gravity 1.0.
Click Finish.
QUICK SUMMARY OF PRESSURE SETUP
For
SURFACE WATER MONITORING
(install probe before
setting these):
Display Mode
Level - Surface
Depth/Level Conversion
Specific Gravity 1.0
Level Reference
Current elevation or gauge
height of water surface
Reference Reading
Now (with probe installed)
For
AQUIFER TESTING:
1 Select Pressure in Parameters list, select Edit
Display Mode
Depth/Level Conversion
Specific Gravity 1.0
2
Display Mode:
Select Level - Top of Casing, click Next.
3 Select Depth/Level Conversion: Specific Gravity 1.0. Click Next.
4 Select Level Reference 0 & Reference Reading Start of Test.
Click Finish.
After setting the pressure mode, select the
UNITS
you want:
7IN3ITU/PTIONSMENU0REFERENCESSCROLLDOWNTO
LEVEL/DEPTH
, select unit, click OK
Pocket-Situ: Tap Home at top of screen, tap Setup at bottom of screen, scroll down to
LEVEL/DEPTH
, select unit, click OK
2
3
4
4
SECTION 7: MONITORING PRESSURE (WATER LEVEL)