![In-situ Aqua Troll 400 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/in-situ/aqua-troll-400/aqua-troll-400_operators-manual_2050333027.webp)
800-446-7488
27
www.in-situ.com
5. Completely submerge the RDO Sensor into the solution.
6. Click
OK
, to start the calibration.
7. When the screen indicates that the calibration has reached stability, click
Accept
to complete the calibration, or
click
Cancel
to return to the previous calibration.
8. You can save or print the calibration report.
9. Click
OK
to complete the calibration.
10. Once calibration is complete, remove the instrument from the calibration cup and rinse both thoroughly with clean
water.