Calibrate and Set Up Sensors
The
Sensors
tab allows you to view the sensors that are available on the instrument. From this tab you can access
calibration Wizards and sensor setup options. You can also view sensor serial numbers, factory calibration dates, and user
calibration dates.
Screen Element
Purpose
The
Calibration
button starts the Calibration Wizard for the
selected sensor.
This button opens the setup options for the selected sensor.
These options include selecting parameters, setting units, and
setting sentinel values.
When you click on the sensor you want to calibrate or configure, the
Calibration
button and the
Sensor Setup
buttons
become active.
Calibration Frequency Recommendations
In-Situ sensors are factory calibrated across the entire range of each sensor, and thus achieve a very high degree of
accuracy and stability for extended periods of time without user calibration. In-Situ recommends inserting the instrument into
a known calibration standard to check the accuracy of a sensor prior to performing any user calibration if you suspect drift,
unless a user calibration is required by a standard operating procedure. Calibration requirements will vary by application and
fouling conditions.