Page 23
English
Using the SmartLung for tests in
accordance with the standard
Ventilators and anaesthesia equipment must be verified
in accordance with the following standards:
•
EN 60601-2-12 Critical care ventilators
(substitute for DIN EN 794-1:2001-02)
•
EN 60601-2-13 Anaesthetic systems
(replacement for DIN EN 740:1999-02)
•
EN 794-3 Transport ventilators
•
ASTM F1100-90 (1997) Critical care ventilators
(US FDA)
These standards determine the following settings for
the SmartLung:
Adult
Paediatric
Infant,
neonatal
Tidal volume
Vt (mL)
500
300
30
Rate (bpm)
10
20
30
I:E ratio
1:2
1:2
1:2
Test lung
used
SmartLung
2000 2 L
SmartLung
2000 2 L
SmartLung
Infant 0.5 L
Compliance
(mL/mbar)
50
20
1
Resistance Rp 5
20
50
SmartLung settings for tests in accordance with the standard