Page 17
English
Preface
The SmartLung 2000, SmartLung Adult and the Smart-
Lung Infant are the same size as conventional test lungs
but offer both an adjustable compliance and a resistan-
ce that can be varied between Rp5 and Rp200. The
adjustable leakage also enables the user to verify the
leak compensation of the ventilator. The patient trigger
functions can also be verified.
The SmartLung is the ideal tool for verifying ventilators
and anaesthesia equipment and is the ideal supple-
ment for CITREX and FlowAnalyser flow and pressure
measurement devices.
Typical settings
SmartLung
2000
2 L
SmartLung
Adult
1 L
SmartLung
Infant
0.5 L
Adult
Compliance
(mL/mbar)
25, 60, 75
10, 15, 20,
30
Compliance
(mL/kPa)
200, 500
Resistance
Rp5, Rp20,
Rp50
Rp5, Rp20,
Rp50
Paediatric
Compliance
(mL/mbar)
10, 15, 20
Resistance
Rp20, Rp50
Infant
Compliance
(mL/mbar)
1, 2, 3, 5
Resistance
Rp50,
Rp200
Technical Data
SmartLung
2000
2 L
SmartLung
Adult
1 L
SmartLung
Infant
0.5 L
Compliance
(mL/mbar)
25, 60, 75
10, 15, 20,
30
1, 2, 3, 5
Resistance
Rp5, Rp20, Rp50, Rp200
Volume (mL)
0 –1000
2 L bag
0 – 600
1 L bag
0 – 200 mL
0.5 L bag
Leak (L/min)
0 – 10
Weight (g)
400
325
285
Dimensions (mm) 400×140×40 303×115×40 270×115×40
Deadspace (mL)
3.7
Components
replaceable
Cleaning
by disinfection
* Tolerances: Resistance ± 10 %, compliance ± 20 %
Содержание SmartLung 2000
Страница 2: ...Deutsch 3 English 17 Espa ol 31...
Страница 12: ...Seite 12 Deutsch Silikonbeutel herausziehen Die Montage erfolgt in umgekehrter Reihenfolge...
Страница 16: ......
Страница 26: ...Page 26 English Pull out silicone bag The lung is assembled in the reverse order...
Страница 30: ......
Страница 40: ...P gina 40 Espa ol Sacar la bolsa de silicona El montaje se realiza en el orden inverso...