23
Alternatively, you can remove the charging wand from
your implant site,
which will cause your
Vesta Charger to
time out and
shut off automatically.
Note:
If you wish to resume charging your OPTIMIZER
Sm
art Mini IPG after
you terminate a charging session,
please wait for approximately 10 minutes before initiating
a new charging session to allow the temperature of your
OPTIMIZER Smart Mini IPG to return to its baseline
temperature.
6.8.2
Termination of Charging Session Due to IPG
Temperature
T
o ensure your safety while charging your
OPTIMIZER
Smart Mini IPG, the temperature of your IPG is monitored
during the charging process. If
the
reported temperature
of
your OPTIMIZER Smart Mini IPG
at the beginning of
the charging session is out
side
the accept
able
temperature range or if the temperature of
your implanted
OPTIMIZER Smart Mini IPG
remaining consistently high
for more than 10 minutes
while it is being charged, then
your Vesta
Charger will
emit 3
long beeping ton
es
and
display the Charging IPG Temperature Error
screen,
indicated by a thermometer icon in the center of the
screen (see
Figure 19
). Your Charger will then shut off.
If
this should occur, please wait for approximately 10
minutes before initiating a new c
harging session.
Figure 19: Charging IPG Temperature Error Screen