20
7.1
Water tight connectors and cable confection for use with the
environmental measurement system ENVIS
Dear Customer,
water tightness of the sensor modules (IP65) and the connectors (IP67) can only be guaranteed by
observing the following:
1. In case of carrying out the cabling of the distribution modules by yourself, please make sure
that the cable seals fit tightly. Only cables with a sheath diameter of 5 to 8mm should be used.
In case of smaller diameters water may enter the connector, the distribution box or the
measurement device and may alter measurements or even destroy the equipment. IMKO
supplies special seals.
2. All unused connectors have to be protected by blind covers. IMKO supplies all equipment
respectively all open connecting terminals protected by blind covers. Additional ones can be
obtained by IMKO.
3. When installing the connectors, special attention has to be paid to the connectors being
threaded with care. The threads have to be aligned correctly to ensure efficient tightness. In
addition, the connector should only be hand-screwed, i.e. without use of tools (e.g. pliers).
A thin coating of grease should be applied to the threads of the threaded ring and to the
connector socket in order to ensure easy removal even after long periods of use.