![IMKO TRIME-FM Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/imko/trime-fm/trime-fm_user-manual_2047647017.webp)
17
S
TEP
L
IST
B
ASIC
A
LIGNMENT
1
Connect the probe (P3 etc.!) to the TRIME-FM.
2
Press START button for switching on the TRIME-FM. Display shows:
TRIME-System
Software Version 2.3
SN XXXX
(=serial number of the TRIME-FM)
3
Connect the calibration connector at the 7pin flange connector of the
TRIME-FM. Display shows:
BASIC CALIBRATION
1. dry glass beads
Reference 2.9Vol%
4
Put the P3 probe into your dry glass beads. (read “Preparation of the
glass beads”p.16)
5
Press START button for starting the measurement
Display shows 1
st
:
BASIC CALIBRATION
1. dry glass beads
Checking TDR Circuit
Display shows 2
nd
:
BASIC CALIBRATION
1. dry glass beads
Measuring.......
Display shows 3
rd
:
BASIC CALIBRATION
2. sat. glass beads
Reference 43.7Vol%
6
Put the P3 probe into your saturated glass beads (Read the description
of the TRIME-FM user manual too!)
7
Press START button for starting the measurement
Display shows 1
st
:
BASIC CALIBRATION
2. sat. glass beads
Checking TDR Circuit
Display shows 2
nd
:
BASIC CALIBRATION
2. sat. glass beads
Measuring.......
Display shows 3
rd
:
BASIC CALIBRATION
1. dry glass beads
Okay, next probe!
Reference 2.9Vol%
6
Probe is calibrated.
You can connect the next P3 probe for calibration (begin with step
1
)
You finished calibration work? Then unplug the calibration connector!
Содержание TRIME-FM
Страница 1: ...TRIME FM User Manual ...