25
9. Error messages
Control Unit U8200, U8300
The error messages are indicated at the rear of the control unit by the status LED (flashing green).
Flashing cycle
Description of error
Solution
1x
Overheating
> Switch off conrol unit
> Allow the control unit to cool for at least 10 minutes
> Observe permissible ambient temperature/operating mode
2x
Foot control
> Release foot control
5x
Time-out (> 15 min)
> Release foot control
(must not be active for longer than 15 minutes without interruption)
6x
Handpiece
> Check tip
(full engagement, damage, torque)
> Dry the handpiece/supply hose
> Check plug-in connection of the handpiece/supply hose
> If the error message appears again, contact an authorised iM3 service
partner immediately.
8x
System error
> Start the device again
> Contact an authorised iM3 service partner.