![iM3 SP6 Скачать руководство пользователя страница 28](http://html1.mh-extra.com/html/im3/sp6/sp6_instructions-for-use-and-technical-description_2044814028.webp)
- 26 -
SECTION X: SPECIFICATIONS
The ART-SP6 scaler is designed and manufactured to meet the most demanding environment. It’s
specifications are listed below
Classification
Protection against electric shock
:
Class I
Degree of protection against electric shock
:
Type BF
Mode of operation
:
Continuous
Medical device directive
:
IIa
Standards Followed
ISO13485:
Medical Devices, Quality Management Systems, Requirements for regulatory
purposes
93/42/EEC:
MDD CE Directive
EN60601-1:
Medical Electrical Equipment, General Requirement for Safety
EN60601-1-2:
Medical Electrical Equipment, Electromagnetic Compatibility,
Requirements and Tests
Specification
•
Power supply
115V ±5%
~
50/60Hz 125VA
230V ±5%
~
50/60Hz 125VA
•
Working frequency
26KHz ~ 32KHz
•
Polisher speed
3000
~
30000 R.P.M
•
Water supply
25
~
60 PSIG(172~414KPa)
•
Dimension
23cm(L) x 21cm(W) x 9cm(H)
•
Weight
3.7 Kg (including the handpiece)
•
Hand piece Cable
250 cm
•
Footswitch Cable
250 cm
Operation environment
•
Temperature
0
℃
~
40
℃
•
Relative Humidity
10% ~ 90% (non-condensing)
Transport and storage conditions
•
Temperature
0
℃~ 60℃
•
Relative Humidity
10% ~ 90% (non-condensing)
•
Atmospheric pressure
860~1060 hPa
SECTION XI: DISPOSAL
Please follow county and state regulations for disposal of the ART-SP6.
Содержание SP6
Страница 29: ...Unité ART SP6 Scaler Polisseuse Combo Manuel d utilisation ...
Страница 58: ...ART SP6 Zahnsteinentfernungs und Poliergerät Betriebsanleitung ...
Страница 86: ... 28 Externer Wasserdruck zu niedrig Prüfen Sie den Wasserdruck dieser sollte 25 60 PSI betragen ...
Страница 97: ... 9 No tire del Cable No doblar No retuerce No agrupar Tire para desconectar ...